MRU Instruments, Inc.
USER MANUAL OPTIMA 7 BIOGAS
13
5
MEASUREMENT PRINCIPAL
The analyzer draws a sample of the flue gases from the duct using a built-in gas pump through the probe is cleaned and
dryed using condensate separator with built-in filter and analyzes the extracted gas with electrochemical sensors.
Draft and temperature are measured at the tip of the sampling probe.
5.1
Gas flow diagram
Position
Description
1
Sampling probe
2
Triple hose
3
Condensate separator
4
Star filter
5
Non return valve
6
Gas pump
7
Sensor chamber
8
O2-Sensor
9
Further electrochemical sensors * (CO, NO2, H2S)
10
CO2/CH4 NDIR Sensor
11
Not connected
12
Pressure sensor *
* optional
Содержание OPTIMA 7 BIOGAS
Страница 1: ...MRU Instruments Inc USER MANUAL OPTIMA7 BIOGAS 1...
Страница 39: ...MRU Instruments Inc USER MANUAL OPTIMA 7 BIOGAS 39 OK This user s decision must be confirmed see chap 16 2...
Страница 58: ...MRU Instruments Inc USER MANUAL OPTIMA 7 BIOGAS 58 16 7 Declaration of conformity...
Страница 59: ...MRU Instruments Inc USER MANUAL OPTIMA 7 BIOGAS 59...
Страница 60: ...MRU Instruments Inc USER MANUAL OPTIMA 7 BIOGAS 60 Technical data change w o notice Date of issue 20140731...