Maintenance and cleaning
11.1.
Cleaning and maintenance
The analyser needs only low maintenance effort for long value
preservation:
CAUTION
Acid from the condensate
Acid burns may result from weakly acidic liquids from the con-
densate.
►
If you come into contact with acid, wash the area imme-
diately using a lot of water.
►
Note the safety data sheet for phosphoric acid (10%).
NOTE
Please note that correct operation of the analyser is only
ensured if the sensors are adjusted regularly.
►
Depending on the intensity of use, the sensors have to
be adjusted / calibrated 1-2 times a year.
After every measurement:
►
Remove the gas sampling tube from the analyser, so that the hose can
dry.
►
Check the filter in the probe head. Replace the filter if necessary.
Occasionally:
►
Clean the probe and the probe hose.
►
If the analyser was not used for a longer period of time, charge the bat-
tery first.
►
Charge the battery approximately every 4 weeks.
►
Check the round filter at the front of the instrument. Replace the filter if
necessary.
11.2.
Service check
An annual service check and if necessary, adjustment of the sensors at an
MRU service department
are recommended for the preserva-
tion of value
.