USER MANUAL
MGAprime
MRU GmbH, D-74172 Neckarsulm
41 / 126
Setting lower / higher limit
NOTE
Setting lower limit (4mA)
This setting determines the lower end value, correspond-
ing to 4 mA. If the measured value falls below the set
value, the Analog output stops at 4 mA.
HINWEIS
Setting higher limit (20mA)
This setting determines the higher end value, correspond-
ing to 20 mA. If the measured value rises above the set
value, the value remains at 20 mA.
►
Open the menu Settings.
See also Chapter 5.1 Open menu Settings, Page 26.
►
Press Analog output setup (4-20mA).
A window appears.
►
Press the 4 mA or 20 mA field to set the limits for a specific channel.
A window appears.
►
Enter the desired values for the lower and higher limit.
►
Press OK.
►
Press "Default" to set the default values for the lower and upper
limits.
The lower and higher limit are assigned to the desired channel.