15.3.
Switching off the analyser in case error
If the touch screen does not respond, you have two options for switching off
the analyser:
►
Touch the
⏻
button during flashing LED for 30 sec. minimum
►
After change to continuous lighting, release the
⏻
button
The instrument will switching off with reset.
Alternatively, you can switch off the analyser as follows:
►
Press the reset button on the right side of the analyser.
See also the illustration in 3.4 Connectors, Page 14.
15.4.
Connecting analyser with MRU4win
The analyser is a Modbus device.
►
You can connect Modbus devices with MRU4win via a serial interface (RS
485).
See also chapter 14.5RS485 Extern, Page.111.
►
You can connect Modbus devices to MRU4win via a TCP/IP
connection.
The TCP/IP connection can be made by LAN or WLAN.
Connecting analyser with MRU4win via serial interface (RS 485).
►
Connect the electrical connections of the RS485 interface as
described in Chapter 14.5RS485 Extern, Page111..
►
Connect the analyser to your PC via the RS485 interface.
►
Open the Menu Device settings.
See also Chapter 5.4 Device settings, Page 28.
►
Set "Modbus Slave ID" to 1.
Please note that the Modbus slave ID is a consecutive number.
If the number 1 is already assigned, assign "Modbus Slave ID" 2 etc. to the
analyser.
►
You can now create the analyser in MRU4win.
See also Chapter Creating analyser in MRU4win, Page 119.