Go to the Extras menu.
►
Select Leak proof test.
►
Press OK.
The menu Leak proof test appears.
►
Ensure that the leak test cap is plugged on.
Pressure is established.
A 10-second test is running.
A message appears whether the leak proof test was passed or
not.
►
Remove the proof test cap.
If the leak proof test is not passed, you must check the probe including
the tubing and the condensate separator.
If no leakage is detected on these external parts, the analyzer has to be
check service center (service centers at www.mru.eu).
9.5.
Contents SD card
►
Go to the Extras menu.
►
Select Contents SD card.
►
Press OK.
The menu Contents SD card appears.
The files stored on the SD card are displayed.
►
If necessary, press F2 to refresh.
►
If necessary, press F3 to open the file.