USER MANUAL AMPRO
plus
MRU GmbH, D-74172 Neckarsulm
29 / 100
5.6.
Setting date and time
If the built-in rechargeable battery is completely discharged, a new setting
of these values is required
►
Press F3.
The Extra menu appears.
►
Select Date / Time.
►
Press OK.
The menu Date/Time appears.
►
Press F2.
Das Menü Uhr stellen erscheint.
The menu Set clock appears.
►
Set the desired Date
►
Set the desired Time.
►
Press F2.
The change is saved.
5.7.Configure measurement program
Select one of the 4 configurable measurement programs.
For each of the programs the following parameters can be configured:
CO ppm limit: adjustable value for the CO sensor protection. If the CO
value in the flue gas is higher than the adjusted value in the analyzer, the
purge pump will be activated andthe sensor will be protected against
high CO concentrations. (Option)
Selectable fuel types: choose and select from the available fuel type list
Measurement windows: configuration of what and where will be dis-
played in the 3 measurement value windows.
Zoom – window: select what will be displayed in the zoom - window
Program name
The following programs pre-configured:
“Test Program” indented for use at instrument maintenance and calibra-
tion
Содержание AMPROplus
Страница 1: ...AMPROplus 4202DE USER MANUAL...
Страница 17: ...USER MANUAL AMPROplus MRU GmbH D 74172 Neckarsulm 17 100 Condensate separator with out water stop filter...
Страница 79: ...USER MANUAL AMPROplus MRU GmbH D 74172 Neckarsulm 79 100...
Страница 99: ...USER MANUAL AMPROplus MRU GmbH D 74172 Neckarsulm 99 100 17 Konformit tserkl rung...