Instructions
Display window
off Platform not
shaking
Platform shaking
but the display
window
malfunctioning
Display window on
but platform not
shaking
High temperature
fluctuation
Loud noise
No power
Burned fuse
High frequency
interference source
exists in the same
ower line
Poor contact of door
switch
Frequent door opening
Unleveled instrument
Loose platform
Loose flask clamp
Plug not connected
Power outlet out of power
Power switch off
Change to a new fuse with the same specification
Position: at the powr
outlet
Position: inside the
electrical cabinet
Remove interference source of the same power
line or use a dedicated power line
Press the confirm button to recover display
Contact local sales/service
Reduce times of door opening and decrease
the time consumed when opening the door
Adjust footings to level the instrument
Fasten the fixing screws of the platform
Fasten the fixing screws of the flask clamp
11. Troubleshooting
25
26
Code
Customer tel.: 972-3-5595252