9. Settings

the temperature parameters may be inaccurate and other unpredictable problems

such as inprecise temperature control or temperature out of control may occur

during normal operation.

d) CO

2

 Settings

Press the “       ” button and enter the CO

2

 concentration settings. Press the “       ” button

to move the cursor to the digit that needs to be modified and press the “        ” or “        ”

button to change the current value. Press the “        ” button after the settings are finished.

e) CO

2

 Calibration

During normal operation, press the “       ” button first and then the “        ” button to enter

the CO

2

 concentration calibration mode. The first line shown on the display window is the

current CO

2

 concentration the sensor measures and the second line is the target value.

Press the “        ” , “        ”, and “      ” buttons to adjust the target value. Press the “        ”

button again and then the “       ” to enter normal operation. The calibrated value is saved

and the CO

2

 calibration is now successful. 

 

If press the “       ” button in the calibration

mode, the instrument will exit the CO

concentration calibration mode, the calibrated value

are not saved.If press the “       ” button in the calibration mode, the instrument will exit the

CO

2

 calibration and the target value is not saved and reset to factory settings.

The instrument will exit the CO

2

 calibration automatically if there is no operation in 10

seconds in the CO

calibration mode and the target value is not saved.

f) Illuminance Setting

During the normal operation, press the “       ” button when “Light Strength: X ” is shown on

the display window to make the cursor flashing at the X point. Press the “       ” and “       ”

buttons to choose the required level and the illuminance setting is finished.

Temp.: 
Speed: 
Light Strength: X
TimeUp: 
Main Interface

37.0

100rpm

000:00:00

10. Alarm System

Alarm prompt:

 

Overspeed alarm interface

Speeding Alarm

Stop and Return

Stop and Return

Stop and Return

The “Overspeed Alarm” line will keep flashing, the buzz will sound intermittently,
and the instrument will stop running.

● 

Overtemperature alarm interface

Overheat Alarm

The “Overtemperature Alarm” line will keep flashing, the buzz will sound intermittently,
and the instrument will stop running.

 

Communication Failure Alarm Interface

Com. Fault Alarm

The “Communication Failure Alarm” line will keep flashing.

 

    The buzz will sound every 2 minutes.

Door Open Alarm 

Содержание IncuShaker LOM-7450-CO2

Страница 1: ...Stackable Incubator Shaker User s Manual Please read this user s manual carefully before using the product LOM 7450 7450 CO2 7450 L...

Страница 2: ...choosing IncuShaker Series Incubator Shaker Your trust and support in MRC LTD are greatly appreciated MRC Incubator Shaker Series is model designed for shaking and heat samples according to the resear...

Страница 3: ...e instruments are running with the maximum load it is recommended to set the rotation speed no more than 250 rpm 11 Please make sure the vent of the instrument is kept away from other vents or heat so...

Страница 4: ...rview 2 2 Features Large view window LCD dot matrix display User friendly interface Multi segment programming Extra long timing up to 999 hours Soft start and smooth acceleration Open door cut off swi...

Страница 5: ...Range Platform Size Platform Number Maximum Capacity Standard Configuration Power Power Supply Overall Size Net Weight CO2 Concentration Rang CO2 Concentration Precision CO2 Detecting sensor Chamber H...

Страница 6: ...X H c Overal size when stacking three units W X D X H 124 78 67cm 124 132 67cm 124 78 200cm Please leave adequate space around the instrument to help the cooling system of the instrument to dissipate...

Страница 7: ...shown in figure 2 It may cause the pipe connect the incubator and the inlet port to fall off and the leak of the CO2 gas if the secondary pressure is too high High CO2 concentration will do harm to yo...

Страница 8: ...tening screws on the flask clamp regularly to prevent it from falling off Please clean the instrument regularly Please check if the v belt at the bottom of the instrument is not loosened after using t...

Страница 9: ...to stop the instrument when it s in Simple Operation or Timing Program mode 9 Settings Turn on the power switch at the right side of the instrument after the instrument is properly placed Thanks for...

Страница 10: ...s after finishing running the program Return to Timing Program page When Timing Program Operation shows on the display window press the button to choose whether to continue running Stoppig Y N Press t...

Страница 11: ...ing Main Interface Press the button to choose a function and enter the button to enter the corresponding operation page a Basic settings Press the button to choose the parameter that needs to be modif...

Страница 12: ...e current parameters and return to the temperature control parameter settings C 2 2 Temperature Calibration The value outside of the bracket is the target temperature value the value inside the bracke...

Страница 13: ...he calibrated value are not saved If press the button in the calibration mode the instrument will exit the CO2 calibration and the target value is not saved and reset to factory settings The instrumen...

Страница 14: ...for more than 5 minutes The alarm indicator light will flash and the buzz will sound The electromagnetic gas valve will be closed automatically Press any key to cancel the alarm status and the instru...

Страница 15: ...t connected Power outlet out of power Power switch off Change to a new fuse with the same specification Position at the powr outlet Position inside the electrical cabinet Remove interference source of...

Страница 16: ...For Model LOM 7450 CO2 only For Model LOM 7450 CO2 only For Model LOM 7450 CO2 only For Model LOM 7450 CO2 only For Model LOM 7450 CO2 only Optional Accessories Name Specifications Name Specifications...

Отзывы: