9. Settings
9. Settings
Press the “ ” button to choose the parameter that needs to be modified or “Main
Interface”; press the “ ” or “ ” button to move the cursor to the digit that needs to
be modified and press the “ ” or “ ” button to change the value; press the “ ”
button to enter the password and enter the “password setting” page if the password
enterted is correct; press the “ ” button to enter password and enter the auto adjust the
temp. parameters page if the password entered is correct.
Note: In order to prevent misoperation, the manufacturer will keep the original password.
If the password is lost, please contact the customer service department of the
manufacturer immediately.
C.1 Password Setting Page
Press the “ ” button to choose the parameter that needs to be modified or “Main
Interface”; press the “ ” or “ ” button to move the cursor to the digit that needs to
be modified and pres the “ ” or “ ” button to change the calue; press the “ ”
button to save the password and return to the instrument settings page (the password is
not saved if the cursor is at back to the instrument settings).
C.2 Temperature Control Parameters Settings
Temp. Display
Temp. Calibrate
Temp. Setting
Return Main Set
C.2.1 Temperature Display
Press the “ ” button to choose the parameter that needs to be modified; press the
“ ” or “ ” button to move the cursor to the digit that needs to be modified and press
the “ ” or “ ” button to modify the value (-5 to 4, resets at -5); press the “ ” button
to save the current parameters and return to the temperature control parameter settings
C.2.2 Temperature Calibration
The value outside of the bracket is the target temperature value, the value inside the
bracket is the difference between the actual temperature and the target temperature; the
emperature values of A, B, C, D, E, F listed in the small to large order are (04.0, 10.0, 20.0,
37.0, 50.0, 60.0). Press the “ ” button to choose the parameter that needs to be
modified; press the “ ” or “ ” button to move the cursor to the digit that needs to be
modified and press the “ ” or “ ” button to change the current value (-9 to +9); press
the “ ” button to save the current parameter settings and return to the temperature
control parameter settings page.
C.2.3 Temperature Settings
It will enter the next page after 2 seconds or if pressing the “ ” button.
Press the “ ” button to exit the temperature control parameters settings.
Choose the temperature control parameter settings to enter the corresponding
settings. The whole process takes about 2 hourse. During this process, the power
should not be cut off and the incubator door is not allowed to be opened. Otherwise
page; T1, T2, T3, T4 are the current temperature values of the 4 temperature sensors
inside the incubator, the value in the bracket is the temperature difference between the
sensor value and the precise temperature detector value; (for example: T1: 37 °C, means
the temperature sensor shows the temperature inside the incubator, if the precise
temperature detector at the same spot of the sensor shows the temperature to be 36 °C,
-1.0 will be shown in the bracket, ie (-1.0)).
Set Password:
XXXX
Return Main Set
Temp.is
auto adjusting
Don’t touch the
pad. About 2hours
Temp.
is Auto Adjusting
Please hold on