9. Settings

9. Settings

Press the “        ” button to choose the parameter that needs to be modified or “Main

Interface”; press the “        ” or “        ” button to move the cursor to the digit that needs to

be modified and press the “        ” or “        ” button to change the value; press the “        ”

button to enter the password and enter the “password setting” page if the password

enterted is correct; press the “        ” button to enter password and enter the auto adjust the 

temp. parameters page if the password entered is correct.

Note: In order to prevent misoperation, the manufacturer will keep the original password.
If the password is lost, please contact the customer service department of the
manufacturer immediately.

C.1 Password Setting Page

Press the “        ” button to choose the parameter that needs to be modified or “Main

Interface”; press the “        ” or “        ” button to move the cursor to the digit that needs to

be modified and pres the “        ” or “        ” button to change the calue; press the “        ” 

button to save the password and return to the instrument settings page (the password is

not saved if the cursor is at back to the instrument settings).

C.2 Temperature Control Parameters Settings

Temp. Display

Temp. Calibrate

Temp. Setting

Return Main Set

C.2.1 Temperature Display

Press the “        ” button to choose the parameter that needs to be modified; press the

“        ” or “        ” button to move the cursor to the digit that needs to be modified and press

the “        ” or “       ” button to modify the value (-5 to 4, resets at -5); press the “       ” button

to save the current parameters and return to the temperature control parameter settings

C.2.2 Temperature Calibration

The value outside of the bracket is the target temperature value, the value inside the

bracket is the difference between the actual temperature and the target temperature; the 

emperature values of A, B, C, D, E, F listed in the small to large order are (04.0, 10.0, 20.0,

 37.0, 50.0, 60.0). Press the “       ” button to choose the parameter that needs to be

modified; press the “       ” or “        ” button to move the cursor to the digit that needs to be

modified and press the “       ” or “       ” button to change the current value (-9 to +9); press

the “       ” button to save the current parameter settings and return to the temperature

control parameter settings page.

C.2.3 Temperature Settings

It will enter the next page after 2 seconds or if pressing the “        ” button.

Press the “       ” button to exit the temperature control parameters settings.

Choose the temperature control parameter settings to enter the corresponding
settings. The whole process takes about 2 hourse. During this process, the power
should not be cut off and the incubator door is not allowed to be opened. Otherwise

page; T1, T2, T3, T4 are the current temperature values of the 4 temperature sensors
inside the incubator, the value in the bracket is the temperature difference between the
sensor value and the precise temperature detector value; (for example: T1: 37 °C, means
the temperature sensor shows the temperature inside the incubator, if the precise
temperature detector at the same spot of the sensor shows the temperature to be 36 °C, 
-1.0 will be shown in the bracket, ie (-1.0)).

Set Password:

XXXX

Return Main Set 

Temp.is
auto adjusting
Don’t touch the
pad. About 2hours

Temp.
is Auto Adjusting
Please hold on 

parameters

Содержание IncuShaker LOM-7450-CO2

Страница 1: ...Stackable Incubator Shaker User s Manual Please read this user s manual carefully before using the product LOM 7450 7450 CO2 7450 L...

Страница 2: ...choosing IncuShaker Series Incubator Shaker Your trust and support in MRC LTD are greatly appreciated MRC Incubator Shaker Series is model designed for shaking and heat samples according to the resear...

Страница 3: ...e instruments are running with the maximum load it is recommended to set the rotation speed no more than 250 rpm 11 Please make sure the vent of the instrument is kept away from other vents or heat so...

Страница 4: ...rview 2 2 Features Large view window LCD dot matrix display User friendly interface Multi segment programming Extra long timing up to 999 hours Soft start and smooth acceleration Open door cut off swi...

Страница 5: ...Range Platform Size Platform Number Maximum Capacity Standard Configuration Power Power Supply Overall Size Net Weight CO2 Concentration Rang CO2 Concentration Precision CO2 Detecting sensor Chamber H...

Страница 6: ...X H c Overal size when stacking three units W X D X H 124 78 67cm 124 132 67cm 124 78 200cm Please leave adequate space around the instrument to help the cooling system of the instrument to dissipate...

Страница 7: ...shown in figure 2 It may cause the pipe connect the incubator and the inlet port to fall off and the leak of the CO2 gas if the secondary pressure is too high High CO2 concentration will do harm to yo...

Страница 8: ...tening screws on the flask clamp regularly to prevent it from falling off Please clean the instrument regularly Please check if the v belt at the bottom of the instrument is not loosened after using t...

Страница 9: ...to stop the instrument when it s in Simple Operation or Timing Program mode 9 Settings Turn on the power switch at the right side of the instrument after the instrument is properly placed Thanks for...

Страница 10: ...s after finishing running the program Return to Timing Program page When Timing Program Operation shows on the display window press the button to choose whether to continue running Stoppig Y N Press t...

Страница 11: ...ing Main Interface Press the button to choose a function and enter the button to enter the corresponding operation page a Basic settings Press the button to choose the parameter that needs to be modif...

Страница 12: ...e current parameters and return to the temperature control parameter settings C 2 2 Temperature Calibration The value outside of the bracket is the target temperature value the value inside the bracke...

Страница 13: ...he calibrated value are not saved If press the button in the calibration mode the instrument will exit the CO2 calibration and the target value is not saved and reset to factory settings The instrumen...

Страница 14: ...for more than 5 minutes The alarm indicator light will flash and the buzz will sound The electromagnetic gas valve will be closed automatically Press any key to cancel the alarm status and the instru...

Страница 15: ...t connected Power outlet out of power Power switch off Change to a new fuse with the same specification Position at the powr outlet Position inside the electrical cabinet Remove interference source of...

Страница 16: ...For Model LOM 7450 CO2 only For Model LOM 7450 CO2 only For Model LOM 7450 CO2 only For Model LOM 7450 CO2 only For Model LOM 7450 CO2 only Optional Accessories Name Specifications Name Specifications...

Отзывы: