2)After Display unit is selected to " C " or " F ", press the
" Enter Button " ( 3-4, Fig. 1 ) will save the setting
function with default.
10-8 Set CD temperature compensation factor
When the lower display show " PEr C "
1)This function only for the Conductivity ( TDS ) mode of
adjusting the probe's Temp.compensation value in
%/per unit. The default value is 2 %/ per .
℃
℃
2)Use the " Button " ( 3-5, Fig. 1 ) or " Button "
▲
▼
( 3-6, Fig. 1 ) to select the upper value to the desired
Temp. compensation value ( %/per ), then press the
℃
" Enter Button " ( 3-4, Fig. 1 ) will save the setting
value temporally.
10-9 ESC
When the display show " ESC "
When the Display show the text " ESC ", then press the
" Enter Button " ( 3-4, Fig. 1 ) will finish the Advanced
Setting procedures and return to the normal measuring
screen.
Remark :
During execute the " Advanced Setting " function,
if press " ESC Button " ( 3-3, Fig. 1 ) will exit the
" Advanced Setting " function, the LCD will return
to normal screen.