6)Press the " Enter Button " ( 3-4, Fig. 1 ), the measuring
value will present on both upper and lower Display.
Use " Button " ( 3-5, Fig. 1 ), " Button " ( 3-6, Fig. 1 )
▲
▼
to adjust the up display value exact same as the standard
conductivity value. Press the " Enter Button " ( 3-4. Fig. 1 )
will save the calibration data and finish the calibration
procedures.
1.318
1.413
Standard value
1.318
1.318
* If only intend to make the one point calibration,
just execute the 2 mS range ( 1.413 mS Cal. )
is enough.
* Multi-points calibration procedures should
execute the 2 mS range ( 1.413 mS Cal. )
calibration at first, then make other ranges
(20 uS range, 20 mS range or 200 mS range )
calibration procedures following if necessary.
12