Optional
* 1.413 mS Conductivity Standard
Accessories
Solution...............................CD-14
SD memory card ( 1 GB )
SD memory card ( 2 GB )
AC to DC 9V adapter.
USB cable, USB-01.
RS232 cable, UPCB-02.
Data Acquisition software,SW-U801-WIN.
2-2 Electrical Specifications (23± 5 )
℃
Conductivity ( uS, mS )
Range
Measurement
ResolutionAccuracy
200 uS
0 to 200.0 uS
0.1 uS
2 mS
0.2 to 2.000 mS
0.001 mS
± (2% F.S.+1d)
20 mS
2 to 20.00 mS
0.01 mS
* F.S. -
200 mS
20 to 200.0 mS
0.1 mS
full scale
* Temperature Compensation :
Automatic from 0 to 60 ( 32 - 140 ), with temperature
℃
℉
compensation factor variable between 0 to 5.0% per C.
* The accuracy is specified under measurement value 100 mS.
≦
* mS - milli Simens
* @ 23± 5
℃
TDS ( Total Dissolved Solids )
Range
Measurement
ResolutionAccuracy
200 PPM
0 to 132 PPM
0.1 PPM
2,000 PPM
132 to 1,320 PPM
1 PPM
± (2% F.S.+1d)
20,000 PPM 1,320 to 13,200 PPM
10 PPM
* F.S. -
200,000 PPM 13,200 to 132,000 PPM
100 PPM
full scale
* Temperature Compensation :
Automatic from 0 to 60 ( 32 - 140 ), with temperature
℃
℉
compensation factor variable between 0 to 5.0% per .
℃
* The accuracy is specified under measurement value 66,000 PPM.
≦
* PPM - parts per million * @ 23± 5
℃