6 Maintenance and Inspection
The user (medical institutions, hospitals and clinics) is responsible for maintenance and inspection of the medical device.
* This apparatus should be inspected every 6 months in accordance with the following maintenance and inspection items.
* Maintenance and inspection are generally considered to be the duty and obligation of the user, but if, for some reason, the user
is unable to carry out these duties, they may rely on a quali
fi
ed medical device service provider. Consult your local dealer or J.
MORITA OFFICE for details.
* If your equipment needs to be repaired, contact your local dealer or J. MORITA OFFICE for details.
• Insert the Tester and check that the indicator is within three bars above or below 1 on the meter.
• Press the Reference button to make sure it changes from 1 to 2 to 3, etc.
• Check that the probe cord can be properly plugged into its jack on the shoulder of the chair.
• Check that the
fi
le holder’s plug can be connected properly to the probe cord and that the
fi
le holder can be clipped onto a
fi
le.
Check the contrary electrode can be plugged into its probe cord connector.
• Touch the contrary electrode with the
fi
le holder and make sure all the bars on the meter light up.
Any medical devices which could possibly be contaminated must be
fi
rst decontaminated by the responsible dentist or doctor and
then be disposed of by an agent licensed and quali
fi
ed to handle medical and industrial waste upon the request from the medical
institution.
Responsibility for Maintenance and Inspection
Items to be inspected
Disposal of Medical Devices
Probe Cord
(1)
Code No. : 3177999
File Holder
(5)
Code No. : 7503670
Contrary Electrode
(5)
Code No. : 7503680
Tester
(1)
Code No. : 9039953