32
4. Usage
Switching Apical Actions
EMR-interlock endodontic mode
In this setting, you can set the motion of the micromotor when the
fi
le tip reaches a self-set position in the root canal (Flash Bar
position).
AS
AR
AS
OAS
(Optimum apical stop)
This function automatically semi-rotates or rotates the micromotor
counterclockwise to resolve
fi
le penetration and stop its motion.
Automatic apical reverse
This function automatically rotates the micromotor counterclockwise.
Automatic apical stop
This function automatically stops the action.
A
Apical action OFF
Turning the action OFF prevents the micromotor from automatically
stopping or rotating counterclockwise so that the operation may
continue even after the
fi
le reaches the Flash Bar position.
* These functions are available only in EMR-interlock endodontic mode.
Setting the Flash Bar Position
EMR-interlock endodontic mode
You can set the position in the root canal where the apical action is being operated.
Set the Flash Bar position (yellow) using “+” or “
−
.”
The position can be set within a range from “Apex” to “2.”
The 0.5 reading on the meter indicates the area near the physiological apical
foramen.
* These functions are available only in EMR-interlock endodontic mode.
10R:1
100 rpm
OFF
OFF
Endo
1
E
OGP
Endo
90
r
AS
OAS
USER A
Page
1
E
USER A
Page
2
Flash Bar position
Area near the physiological apical foramen