28
4. Usage
Optimize rotation control in order to improve torque detection accuracy during endodontic treatment.
EMR-interlock endodontic mode
Page
2
4.4.7 Calibration
* These functions are available only in EMR-interlock endodontic mode.
●
You should calibrate the micromotor:
- Before using it for the
fi
rst time after purchasing.
- After replacing your handpiece with a new one.
- Before using a handpiece di
ff
erent from the one which has been calibrated.
- When the micromotor fails to rotate continuously in the OTR mode and alternates between clockwise rotation and counterclockwise
rotation.
Perform calibration after a speed-down contra
angle handpiece for endodontic treatment has
been attached to the micromotor.
When performing calibration with the
fi
le
attached, prevent your
fi
ngers from touching the
fi
le during rotation.
Calibration in progress.
Calibration has successfully completed.
Calibration has failed.
If calibration has failed, press the Enter key to return to the basic screen to
execute calibration again.
E
USER A
E
CALIBRATION
࢚ ࣥ ࢻ ⏝ ῶ ㏿ ࢥ ࣥ ࢺ ࣛ ࢆ ╔ ࡋ ࡓ
≧ ែ ࡛ ᐇ ⾜ ࡋ ࡚ ࡃ ࡔ ࡉ ࠸ ࠋ
࢟ ࣕ ࣜ ࣈ ࣞ ࣮ ࢩ ࣙ ࣥ ୰ ࡣ ⮬ ື ⓗ
U S P
ࡽ
U S P
ࡲ ࡛
ື స ࡋ ࡲ ࡍ ࠋ
USER A
1
Press the calibration key.
Read tattention screenthoroughly.
2
Press the Enter key
Start calibration.
E
CALIBRATING
USER A
E
CALIBRATION SUCCEEDED
USER A
E
CALIBRATION FAILED
USER A