43
5 Maintenance
• Do not sterilize the components by any method other than autoclaving.
• The
fi
le holder and contrary electrode are extremely hot after autoclaving; do not touch them
until they cool o
ff
.
Autoclaving and drying temperatures must never exceed 135°C (275°F).
For sterilizing
fi
les, follow the manufacturer’s instructions.
4
Sterilization
Autoclave the components after use for each patient.
Recommended temperature and time: 135°C (275°F), six minutes minimum in
a sterilization pouch.
5.1.2 Wiping with Ethanol for Disinfection (Ethanol 70 vol% to 80 vol%)
Wipe the components with a piece of gauze that has been dampened with ethanol for
disinfection (ethanol 70 vol% to 80 vol%) and wrung out thoroughly.
5.4 Transport and Storage
Transport and storage environments:
Temperature: -10°C to 70°C (14°F to 158°F)
Humidity: 10% to 85% (without condensatin)
Atmospheric Pressure: 70kPa to 106kPa
• Avoid frequent or continuous exposure to direct sunlight.
• If the equipment has not been used for a while, make sure it operates properly and safely before using it again.
Never wipe components with any solution other than ethanol for disinfection (ethanol 70 vol% to
80 vol%). Other solutions could cause cracking and discoloration.
Avoid spilling chemical solutions used for treatment on the any of the components. Wipe up
any chemical spills immediately. These chemicals could damage, deform or discolor plastic and
metal. Use extra caution to avoid spilling formalin cresol (FC) and sodium hypochlorite as they
are quite strong.
5.2 Micromotor
5.3 Ultrasonic Scaler
For maintenance of the micromotor, follow the package insert and operation instructions.
For maintenance of the ultrasonic scaler, follow the package insert and operation instructions.