10
ENGLISH
9.3.6 Alarm Mute
Press alarm mute button to temporary suspend the low-pressure alarm (LED light and buzzer).
Should the situation not resolved within 5 mins, the alarm shall resume to notify the patient.
A. PFA (Power Failure Alarm)
During power failure situation, the Power Failure LED light will lit on with buzzer. By pressing the
mute button to disable both buzzer and LED.
B. Low Pressure Indicator
When low pressure LED lights up, the pressure inside of air mattress is below normal. Please refer
to troubleshooting.
C. Tech. Support
This feature will light during mechanical failure situation. User can notify the technician for repair.
9.3.7 Panel Lock
If the control panel remains untouched for 5 minutes, a green LED will light and the control panel
will lock all settings. This will prevent the settings from being accidently changed during normal
operation. To unlock, simply press the Panel lock button for three seconds.
10. INSTALLATION
Unpack the box and check the package contents for completeness.
Package content List:
- Mattress unit x 1 (may not be included if only pump unit is purchased)
- Pump Unit x 1
- User Manual x 1
Inspect equipment for damage which may have occurred during shipment. If there is damage, please
contact your dealer immediately