17
ENGLISH
18. SPARE PARTS/ACCESSORIES
For spare parts and accessories refer to Moretti main catalogue.
19. TROUBLESHOOTING
If your questions can’t be answered with above information, please contact your local agent di-
rectly. They might require a technician to take care of the problem.
PROBLEM
SOLUTION
Power is not ON
• Check if the plug is connected to mains
Alarm is ON
• Check if the power is suddenly shut down.
• Check if the CPR is sealed.
• Check if the connection between air tube connector
to pump unit is tightly secured.
• Check if all tubing connections along mattress are secured
Patient is bottoming out
• Pressure setting might be inadequate for the patient,
adjust comfort range 1 to 2 levels higher and
wait for a few minutes for best comfort.
The mattress doesn’t inflate
• Check if the connection between mattress and pump
unit is tightly secured
• Check if there is any leakage on air cells
Mattress form is loose
• Check if all the snap buttons or straps of mattress are
all securely fastened
• Check if the mattress is fixed to the bed frame by straps
No air produced from
some air outlets of the
air tube connector
• This is normal since there is alternating mode.
Air outlets take turns to produce air during their cycle time
20. GUIDANCE AND MANUFACTURER’S DECLARATION
20.1 Guidance and Manufacturer’s Declaration - Electromagnetic Emissions
This device is intended for use in the electromagnetic environment specified below. The user of this
device should make sure it is used in such an environment.
EMISSIONS TEST
COMPLIANCE
ELECTROMAGNETIC ENVIRONMENT - GUIDANCE
RF emissions CISPR 11
Group 1
The device uses RF energy only for its internal function. Therefore, its
RF emissions are very low and are not likely to cause any interference in
nearby electronic equipment
RF emissions CISPR 11
Class B
The device is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage
power supply network
Harmonic emissions
IEC61000-3-2
Class A
Voltage fluctuations / Flicker
emissions IEC61000-3-3
Complies
Warning:
1. The device should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the device should be observed to
verify normal operation in the configuration in which it will be used.
2.Use of accessories, transducers and cables other than those specified or provided by the manufacturer of this equipment could result in increased
electromagnetic emissions or decreased electromagnetic immunity of this equipment and result in improper operation.
3. Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12
inches) to any part of the pump, including cables specified by the manufacturer. Otherwise, degradation of the performance of this equipment could result.