caregiver insert palm facing up and place between patient’s buttock’s area and static cell. Should
the palm able to touch patient’s buttock, increase the pressure until the skin is least 1/2” inch away
from the palm. Always leave at least 1” inch space between patient’s buttock areas to the static cell
under to prevent bottoming out.
9.3.2. Mode (LAD680)
A. Max Firm
It will be forced to execute every time the power switch is turned on. This insures the pump is
able to reach its maximum operating pressure. Once the max pressure level is reached, the pump
will automatically switch into alternating mode. User can also use this function as full mattress
inflation while ingress/egress the patient for better support. There is a 20 min auto-recovery
mode back to alternating cycle to ensure patient always getting best therapy possible.
B. Alternate
In Alternate therapy mode, the mattress system will alternate every 10 minutes.
C. Static
Press the button to suspend alternating function, if needed. The pressure inside of air tubes will
be adjusted to the same softness. Press the button again; it will switch back to alternating mode.
Under the static mode, cell pressure level will be lowered compare to the same pressure level
from alternating mode.
D. Seat inflate
The seat inflation increases the pressure for every cells to support the patient during upright
position without bottoming out. User can select this additional feature under either static or
alternate mode.
LAD 690
9.3.3 Pressure Adjust (LAD690)
The function can automatically adjust the mattress pressure based on the patient’s weight. To re-
mind user the process is on-going, the function LED lights will be periodically flashed from left to ri-
ght until the Ready LED turn on permanently (the process is completed). There are three conditions