18 1/3/02
3 lnstallation and Setup
4. Connection of supply voltage to the X-3000ce-UW controller
Connect cable between machine main panel and X-3000ce-UW controller.
Select a supply voltage (230/115 VAC) that is turned on when the power to the
machine main panel is turned on. Then:
a)
Turn off machine main panel.
b)
Connect cable in the machine main panel.
c)
Remove fuse from the fuse holder on the printed circuit board (PCB) if fuse
is already installed.
d)
Connect cable in the X-3000ce-UW (see diagram below).
e)
Turn on machine main panel again.
f)
Using voltmeter, check connection of terminal points L, N and PE before
inserting fuse.
g)
Turn machine off.
h)
Insert correct fuse and turn machine on.
L N PE
L N PE
220
230
11
5
10 MM Max.
R1 R2
R3
R4
R5
R6
Enter
-
+
Back Space
Down
Up
DO 4
DO 3
DO 2
DO 1
DI 8
DI 7
DI 6
DI 5
DI 4
DI 3
DI 2
DI 1
0
0
1
2
3
4
5
6
7
8
9
1
0
11
1
2
13
1
4
1
5
16
1
7
1
8
19
2
0
2
1
22
2
3
2
4
25
2
6
27
2
8
29
44
4
5
46
4
7
4
8
49
5
0
51
5
2
5
3
54
5
5
56
5
7
58
L N PE
Type : xxxxxxxxxx
Serial no : xxxxxxxx
Voltage : xxxxxxxxx
Current : xxxxxxx
230
11
5
30
3
1
32
3
3 3
4 3
5
36
37
38
39
40
41
42
4
3
11
12
13
1
4
Voltage Switch ( In 230 v Position )
Fuse Holder
FUSE SIZE:
115V -160mA T (slow blowing type) Dia. 5 - 20 mm
230V - 80 mA T (slow blowing type) Dia. 5- 20 mm
WD022.dsf