Programming
EPM1000 User’s Guide
31
Communications Error
Set the EPM1000 to echo back invalid commands.
Command Form
COMMunication ERROR <ON|OFF>
Examples
Input
Response
comm error on
none
syst opt bak off
Unrecognized SCPI command:
syst opt bak off
Display Current Reading
Query the EPM1000 for the data currently displayed on the LCD.
Command Form
CHannel QUERY
Examples
Input
Response
ch query
CHAN: QUERY: 227.4 mV
Identification
Query the EPM1000 for identifying information, such as firmware
revision date, model number, etc.
Command Form
*IDN?
Examples
Input
Response
*idn?
Molectron Detector, Inc.
EPM1000 Ver G3.18 26-Aug-96
†
† Actual reading will vary, depending on installed firmware.
IEEE-488 Address
Set or query the IEEE-488 address of the EPM1000. The EPM1000
must be reset for this command to take effect.
Command Form
COMMunication GPIB ADDRess <?|0..31)
Examples
Input
Response
comm gpib addr 12
none
comm gpib addr ?
COMM: GPIB: ADDRESS 12
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com