Operation
EPM1000 User’s Guide
22
Average Mode
(Continued)
Average Mode will function while measuring Joules, Volts, or Watts, but
changing measurement units will reset the EPM1000 to Normal Mode.
Press to enter Average Mode. The LCD will display a countdown
as the EPM1000 counts its first batch, but thereafter will only display
calculated means. The
annunciator will blink while the EPM1000 is
in Average Mode. Press again to exit Average Mode.
Statistics Mode
Use Statistics Mode to measure statistical qualities of a predetermined
number of samples. This number, the Statistics Batch Size, can be 0 or
any number from 2 to 9999. The EPM1000 can be programmed to stop
after taking a single batch (Manual Restart Mode), or continue sampling a
new statistical batch (Automatic Restart Mode).
Statistics Mode will function while measuring Joules, Volts, or Watts, but
changing units during a batch will reset the unit to Normal Mode.
A Statistics Batch Size of 0 sets the EPM1000 to continuously collect
data, comparing values on the fly and updating the Minimum and
Maximum values. During this continuous collecting, the EPM1000 does
not collect or display the Arithmetical Mean or the Standard Deviation.
Press to enter Statistics Mode. Use the or keys to
sequence through the four statistical values. LCD annunciators indicate
which value is displayed,
for the Mathematical Mean (or Average)
of the batch,
for the Minimum value,
for the Maximum value,
and
for the Standard Deviation. Press again to exit Statistics
Mode.
Set the EPM1000 to Normal Mode before changing any settings.
Audible Beeper
The EPM1000 generates an audible tone during certain error conditions
such as an input overrange or when the thermopile temperature exceeds
100ºC.
To turn the Beeper option on, press .
To turn the Beeper off, press .