Instructions for use
LSI_K_np_en 1206_mmm_V2.08_B2V
33
11.14
ADjUSTMENT
TO
PREVENT
CULTIVATING
MEDIA
AND
TISSUE
CULTURES
DRYING
This mechanical and programme adjustment
reduces markedly the cultivating media and
tissue cultures drying during the use of Incucell
with ventilator. In principle, chamber tightness is
increased. Air inlet flaps cannot be opened. The
chamber contains a moistening plate. Ventilator
is switched off at the door opening.
11.15
IC
AND
ICV
WITH
DECONTAMINATING
FUNCTION
applies to volumes 55, 111, and 222 litres.
The model is designed to be used if there is a
danger of potential chamber contamination by
microorganmisms.
Follow these operation instructions:
a) Switch off the device working in common
mode by means of oFF button.
Insert the So card in the chip card reader,
switch the device on by means of oN button
(the device passes to the decontamination
mode – “Decontamination!” is flashing on
the display). Enter the Service 07 (protective
thermostat setting) in a standard way. Set the
limit temperature (optional value) and confirm
the setting.
b) remove the chip card, leave the service
mode, set the decontamination temperature/
time combination optionally in programme
P1 – P6, and start the decontamination. you
can leave the decontamination mode at any
time by means of oFF button; you enter the
common IC/ICV mode.
If the decontamination temperature drops for any
reason below the lower limit set in Service 41,
the subsequent exposure duration at the preset
temperature is extended by the time spent under
lower temperature.
The decontamination can be documented either
through a printer or SW WarmComm.
11.16
VENTICELL
WITH
STERILIzATION
FUNCTION
Programmes P1, P2, P5 and P6 remain
unchanged. P3 and P4 are replaced by
sterilization programmes SS and SP.
11.16.1
bASIC
STERILIzATION
PROGRAMME
SS
SS programme is the basic sterilization
programme. after you select this programme in
VENTICELL with a sterilization function, letters
“SS” are displayed on the first and second
position.
The device starts after the delayed start ends
(max. 9,999 hours) and works under a preset
temperature. Duration of the preset temperature
exposure can either be limited or unlimited.
11.16.1.1
SS
PROGRAMME
SETTING
Set the cursor to the Programme position
by means of ◄► buttons and switch to SS
programme.
Using ◄► buttons, place the cursor to the graph
and check whether the segment 1 in the graph
is activated (if necessary, change it by means of
▲).
Set the delay time between 00 h 00 min and
9,999 hours in the
fields.
Using ◄► buttons, go to the graph and switch
to the second graph segment _/¯ by means of
▲ button. Set the temperature in °C fields. Move
the cursor to position and set the procedure
duration between 00 h 00 min and 9,999 hours,
or ∞:∞ (set the infinity by setting 9,999 hours,
moving the cursor to the figure 9 in units and
pressing ▲).
11.16.1.2
LCD
-
READ
OUT
SHOWN
11.16.1.3
SS
PROGRAMME
START
Press
STArT/STOp
. you will hear a sound
signal and the propeller symbol in the first
character starts rotating clockwise. See the
display below.
SS =
Program Segment
°C
RH
5 6
S S
/
0 0 . 0
0 0
0 0
:
0
0
1
=