requirements: Common PC with an operational
software Windows 95 and higher. one free port
rS 232 (Com) is required for each connected
device. The maximum length of the connecting
cable is 15 m. For more detailed information
about HW, ask the seller.
11.7
hEpA FiLTEr
HEPA filter is either in a normal version or in the
overpressure version.
The overpressure HEPA filter is equipped with
the external fan, which forces the surrounding air
through the HEPA filter into the chamber and thus
it ensures the overpressure in the chamber as
opposed to the surrounding environment by ca.
20 Pa. In case of operation of the overpressure
HEPA filter there may happen to deterioration of
parameters of the thermal homogeneity.
11.7.1
PLACING
THE
HEPA
FILTER
,
FILTER
DIMENSIONS
,
FUNCTION
AND
FILTRATION
CHARACTERISTIC
OF
THE
AIR
FILTER
Fig. 3 (dimensions are in mm)
Type →
Dimens.
↓
55
111
222
404
707
a (mm)
85
85
85
135
135
B (mm)
290
470
710
1470 1470
C (mm)
50
190
190
190
590
Height
a+B+305
(mm)
680
860
1100 1910 1910
Note: The minimum distance from the filter to the wall
is 50 mm.
11.4
LEFT
DOOR
The lock is placed on the upper part of door‘s
surface near the closing mechanism.
11.5
INDEPENDENT
SENSOR
PT
100
The independent flexible sensor is used for
the material temperature measurement in the
chamber. Up to four sensors can be connected.
The information about the given sensors
temperature is displayed on the position of
time information. Temperature and time are
displayed in turns and the individual sensors
are differentiated from each other by means of
letters.
When printing on the printer, each record
contains more lines. The first line, starting
with figure 1, displays the temperature on the
regulatory sensor. Subsequent lines, starting with
letters, display the flexible sensors temperature.
11.6
SUPPORTIVE
SW
FOR
PC
11.6.1
RECORDING
SW
-
WARMCOMM
-
FOR
PC
UNDER
WINDOWS
The program WarmComm is designed to record
the temperature course in the ovens/incubators.
Data obtained during the regulation are displayed
in a diagram (with time on the horizontal axis and
measured data on the vertical axis).
The program enables to follow the regulation on-
line, to store the regulation course to a file on a
disc and to view the already stored files, to send
e-mails in case of not receiving data from the
cabinet and to send reports of excessive values
being reached within the specified time interval.
The instructions for the programme use are
supplied with the installation programme.
Hardware requirements: Common PC with an
operational software Windows 2000, XP and
higher. one free port rS 232 (Com) is required
for each connected device. The maximum length
of the connecting cable is 15 m.
11.6.2
RECORDING
PRINTING
SW
-
PRINTER
ARCHIVE
-
FOR
PC
UNDER
WINDOWS
The Printer archive programme serves for text
recording from the device print outlet. It serves
as a direct substitution of a physical printer.
The data are recorded to a file on a PC disc
and the programme offers more possibilities of
the recorded data handling including archiving
or printing on a table PC printer. Hardware
Instructions for use
28