device continues automatically from the point
before interruption.
4) If the device is switched off by the button
STArT/STOp
or
ON/OFF
, the programme
starts running from the beginning after
reswitching-on.
Ventilator operation after a programme end:
The operation can be set in Service 06.
Immediate ventilator switch-off after programme
termination is pre-set in all ventilator types.
4.7
PROGRAM
P
6
4.7.1
DESCRIPTION
OF
P
6
Programme P6 works with segments whose
duration is determined by real time and calendar
date.
1. Programme P6 consists of 40 segments
and is determined strictly by real time, i.e. it
follows the preset programme independently
of the switch-on time, power failures, restarts
after an error, etc.
2. Segment 0 (so called “delayed start”) is
compulsory; time and date must be set in it.
It may even be set to the past and means the
“beginning of the programme”. Segments 1 to
39 have the same properties. Each of them
is determined by the end temperature, time
or speed of temperature rise, and ventilator
speed.
3. If cycling is switched on, segment end times
are shifted after each cycle end (forward
by the programme duration). Updated
parameters are thus displayed when you
check the parameters during the device
operation (after
X/W
pressing).
4. The whole P6 programme (all segments at
once) can be shifted by service 64 by any
time interval to both the future and past and
there is no need to rewrite all segments
before a new programme start.
H: XXX
the light of the set segment
I: XX . XX
the month and day at the moment (at P5
impossible to be set )
diSpLAy - rUNNiNg OF ThE prOgrAm
A B C
D
E
G
H
F
a:
∗
5
the rotating symbol and the program
number
B: X
the number of the relevant segment
C: X
the type of the set segment:
> not the last, the end length - see E
>Ι I the last, the end length - see E
∞ the last, infinite length
D: XX . X
the temperature in the chamber at the
moment
E: XX : XX
the segment time at the moment -
hours, minutes
all segments - counting down from the
set value towards zero,
in case of an infinite length of the last
segment counting up from zero
in case the time is higher than 9999 hr.
>>:>> is indicated
F: XXX
the fan speed at the moment
G: XX
not used
H: XXX
not used
Programme P5 continuation after preset
parameters change or programme interruption:
1) The programme is running e.g. in segment 5.
you want to change the setting in a passed
already segment (e.g. segment 4) or in a
future segment (e.g. segment 6). Press
X/W
and change the setting according to
your needs. meanwhile, the programme
will continue running in segment 5. Wait for
several second. The programme returns from
the setting mode to the working segment 5
and continues working.
2) The programme is running e.g. in segment 5.
you want to change the setting of the running
programme 5. Press
X/W
and change the
programme 5 setting according to your needs.
Wait for several seconds. The programme
returns from the setting mode to the beginning
of segment 5 and starts running.
3) The programme is running. Power supply is
interrupted either by switching the main device
switch off or due to power supply system
failure. after the power supply is renewed, the
LSI_K_np_en 1206_mmm_V2.08_B2V
17
Instructions for use
T
tn=max 15
t7
t6
t5
t4
t3
t2
t1
t0
T7
T4
T3
T5 T6
T1 T2
S0
S1
S2
S3
S4
S5
S6
S7
Sn
t
tn=max. 39