System Two LED Installation and Service Manual (1003385 Rev. C)
Page 30
Section 4
Mounting the L17 and LCH Spring Arm
Figure 8: Mounting the L17 and LCH Spring Arms
Two people are required to mount the Spring Arm.
An equipment lift is highly recommended to avoid
injury in lifting the Spring Arm into place.
1. Pull the 2 latch pins
①
upwards and route the 2
securing segments
②
out of the Extension Arm
until they snap into place
⑤
.
2. Insert the Spring Arm pivot
④
straight into the
Extension Arm up to the end stop without twisting.
The plug couplings mounted inside the Extension
Arm and Spring Arm are connected during this
installation step.
3. Insert 2 securing segments
②
into the Extension
Arm up to the end stop such that they snap into
place:
The 2 securing segments
②
must protrude into the
groove
③
on the Spring Arm pivot
④
The 2 spring-actuated latch pins
①
audibly click
into place.
*** Risk of the Light Head dropping ***
If the 2 securing segments
②
are not properly
mounted, there is a risk of the Spring Arm with the Light
Head dropping and causing serious injury. Check the 2
securing segments
②
are securely in place:
The Spring Arm must be inserted into the
Extension Arm up to the end stop.
The 2 securing segments
②
must be inserted into
the Extension up to the end stop such that they
protrude into the groove
③
on the Spring Arm pivot
④
.
The 2 latch pins
①
must audibly snap into place
and be on the same height after the installation.