4
Preparing the Stabil-Ion System for Operation
4-10
Series 370 Stabil-Ion Instruction Manual - 370119
4.
Press the
SELECT
key to see the pressure range display.
5.
Press the
▲
or
▼
arrow keys until you see the scale factor
display.
6.
Press the
SELECT
key so that the scale factor value is flashing.
7.
Press the
▲
or
▼
arrow keys to select the desired value.
8.
Press the
SELECT
key to store the new setting in memory.
9.
Press
CANCEL
to return to the control menu.
(Or press
CANCEL
again to return to the status display for the gauge tube
that is turned ON.)
10.
Repeat steps for IG2. IG2 may be set for a different gas scale factor than IG1.
4.3.3.4 Selecting Gas Type for the
Stabil-Ion Gauge
The Stabil-Ion Gauge comes with a Memory Module containing calibration data for nitrogen. These
data are loaded into the Controller memory as gas “
a
” during the Memory Module loading
procedure. See Section 4.3 on page 4-2. If a Memory Module is purchased and installed for a
second gas “
b
”, gas selection may be done using the following process:
Note: If the Stabil-Ion Gauge calibration is for a gas type other than N
2
(or air), we suggest placing a
label on the slip-in gas card display line indicating the gas type or types used for calibration to
prevent mix-ups (see Figure 4-5 on page 4-11). Labels for common gases are provided as a part of
the standard installation hardware. (See Figure 1-3 on page 1-2.)
The default factory setting for this function is N
2
.
1.
Press the
▼
arrow key until you see the control menu in the display.
2.
Press the
SELECT
key to see the pressure range display.
3.
Press the
▲
or
▼
arrow key until you see the IG gas type display.
4.
Press the
SELECT
key so that the gas type is flashing.
5.
Press the
▲
or
▼
arrow key to select
GAS
type. (Note: this function is only
applicable when the controller contains a factory supplied second gas curve
table that is not a factory default.)
IG1
HIGH
PRESS.
RNG
IG1
LOW
PRESS.
RNG
or
IG1
SCAL
1.000
IG1
SCAL
1.000
IG1
SCAL
.775
IG1
SCAL
.775
Control
Menu
Control
Menu
IG1
HIGH
PRESS.
RNG
IG1
LOW
PRESS.
RNG
or
IG1 GAS a
N2
IG1 GAS a
N2
IG1 GAS b
Ar
Содержание Stabil-Ion 370 Series
Страница 2: ......
Страница 8: ...Table of Contents iv Series 370 Stabil Ion Instruction Manual 370119...
Страница 36: ...2 Initial Setup Procedures 2 16 Series 370 Stabil Ion Instruction Manual 370119...
Страница 46: ...3 Installation 3 10 Series 370 Stabil Ion Instruction Manual 370119...
Страница 58: ...3 Installation 3 22 Series 370 Stabil Ion Instruction Manual 370119...
Страница 121: ......