PM 250 

        SG 1024

ALARM TROUBLESHOOTING *

Note*: the last alarm occured is permanently stored by the CPU, even in absence of main line, and 

can be verified using the Front Panel software (vers. 7.50 and over)

Code

Description

Cause/Solution

0001

Overtemperature 

Alarm

The temperature of the water at output exceeds the max temperature 

presetted via the specific parameter (Tout > Tmax)

a) Check water flow

b) Verify correct setting of Tmax

0010

Reverse Power 

Alarm

The reflected power has reached the alarm threshold for a time longer 

than 20 seconds. Set higher threshold or start the automatic tuning to 

deacrese the reflected power.

Full reverse power condition triggers the alarm in one second.

0011

RF Input  

Malfunction (PLL)

Alarm linked to internal frequency generator: the generated driving 
signal is out of expected value. The CPU turns off the power module 
automatically. After the reset operation, the CPU turns on the power 
module and checks possible alarms. If the alarm persist, contact the 
service.

0100

AC/DC Power 

Supply Alarm

The internal power supply has an output < 10V or it exceeds of +/- 5% 
the expected value after 4 seconds.

0101

Overcurrent Final 

Amplifier (M1-4)

The final power amplifier has reached the limit current. Try to reset and 

restart at lower power. If the alarm persists, contact the service.

1001

Interlock Alarm

Check interlock input (0V = Alarm, +24V= OK)

1010

Driver Alarm

The current of the driver (pre-ampli) exceeds 10 A. Try to reset and 
restart at lower power. If the alarm persists, contact the service.

1011

Communication 

interrupted

Alarm engaged only when the SG is run throught the USB port and no 
valid character is received within time out

Содержание SG 1024

Страница 1: ...SOLID STATE MICROWAVE GENERATOR 1000 W 2450 MHZ TECHNICAL NOTE issue April 2017 SG 1024...

Страница 2: ...Cover Rear...

Страница 3: ...tati membri relative alla compatibilit elettromagnetica Conformity to the Directives is assured through the application of the following standards La conformit alle Direttve assicurata tramite l appli...

Страница 4: ...ies products covered by the Directive as a matter of fact it does not perform a function alone must be included into a larger process equipment and it s for stationary use Nevertheless we invite the U...

Страница 5: ...Industrial scientific and medical ISM radio frequency equipment Electromagnetic disturbance characteristics Limits and methods of measurement With reference to Basic Standard EN55011 this equipment is...

Страница 6: ...e The SG 1024 is offered in two versions the Display version with front panel manual commands and PLC interface to control by I O it carries also one USB standard port which allows to drive read data...

Страница 7: ...ency 26 Output power range Wrf 50 1000 Adjustable range 2 100 Max value differs at frequency band limit Output linearity Wrf 20 Forward power Displayed value vs monitor signal Output stability 1 For i...

Страница 8: ...uction for any doubt then power on the unit through your external contactor Once switched OFF wait at least two seconds before setting it ON again The OFF state between two consecutive ON state must l...

Страница 9: ...Troubleshooting table at page 15 for the complete alarm list and relative codes Note the last alarm occured is permanently stored by the CPU even in absence of main line and can be verified using the...

Страница 10: ...be set in external mode using a signal AUTOMATIC TUNING is a feature of the unit based on a patented algorithm US patent 6 020 794 Due to the possibility to change the operating frequency the user can...

Страница 11: ...pages The number enclosed in brackets refers to the small number associated to names of pages in the previous layout 0 1 At POWER ON the unit shows the Logo its name and software version for a moment...

Страница 12: ...power and parts of the circuit on the final power combiner as well as the coaxial socket may be damaged 6 INFO this page shows the last alarm that occured the software version the generator is runnin...

Страница 13: ...4810 1 Preferred size other length on request max 2 m Coax Cable 7 8 with 7 16 Plugs l 1m 441384811 1 Other length on request max 4 m 3 Transition 7 16 to WR340 MWR26 COAX7 16L Aluminum construction F...

Страница 14: ...nitor signals in the range 0 10Vdc pin 2 and pin 3 ground on pin 15 and 16 respectively Forward and Reverse Power Monitor CONN 2 LOCATION Forward Power Wrf 1 Vmon 100 W TBD Pin no 2 return on pin 15 R...

Страница 15: ...m section or ask our Application Note for the protocol implemented Each SG 1024 generator must be associated to an address the two digital rotating switches must be used the left switch sets the tens...

Страница 16: ...nector MOCPC03 2060362 6 1 CPU Board 800000536 B 7 1 DC Power Supply Board 800000538 8 1 Display Board 800000537 9 1 Fan VE8414NGH 10 1 Line breaker with lever IGM22061192003M1 11 1 Lever 12 1 Solid s...

Страница 17: ...Input Malfunction PLL Alarm linked to internal frequency generator the generated driving signal is out of expected value The CPU turns off the power module automatically After the reset operation the...

Страница 18: ...x Plug right angle 7 16 M F p n MWCOAX7 16L Coax cable length 1 m 1 2 p n 441384810 1M0 Coax cable length 1 m 7 8 p n 441384811 1M0 Cable set for input line and signals only p n 441384524 Waveguide to...

Отзывы: