PR4000F
Chapter 4: Operation
27
SCL
(only
SLAVE
or
EXTRN
)
:
The value of
SCL
(Scaler) determines the ratio of the setpoint which will be applied to the slave.
Example for two mass flow controllers:
Channel 1: MFC 500 sccm full scale, Mode
Independent
. This is the master channel!
Channel 2: MFC 200 sccm full scale, Mode
Slave
.
Scaler
SCL
set to 080.00 sccm.
When the master’s flow readout ranges from 0 –500 sccm the setpoint applied to the slave
ranges proportionally from 0 – 80 sccm ! The gas mixing ratio therefore is set to:
500 : 80 = 6,25 : 1
Example as before but for external setpoint mode
EXTRN
:
Channel 1: MFC 500 sccm full scale, Mode
EXTRN
(this makes channel 1 to be the „slave“ of the
external setpoint source!
Scaler
SCL
set to 350.00 sccm.
This means that the external setpoint applied to channel 1 will be scaled down by a factor of
350 : 500 = 0,7.
Depending on configuration and operational mode it is possible to have the setpoint input(s)
displayed. This however is possible only in the
INDEP
mode.
For optimum control performance it is advised not to set the scaler to values below 20 % off the
MFC. It is possible but you should instead use MFC’s with lower ranges.
Note
The setting of the scaler does not affect the independence mode
INDEP
!
Содержание PR 4000 F
Страница 8: ...List of Figures and Tables PR4000F viii This page left blank...
Страница 12: ...Safety information PR4000F 4 This page left blank...
Страница 19: ...PR4000F Chapter 2 Installation 11 Dimensions Figure 1 Dimensions above Front and Rear Panel below Side View...
Страница 50: ...Chapter 4 Operation PR4000F 42 This page left blank...
Страница 64: ...Appendix A Specifications PR4000F 56 This page left blank...
Страница 66: ...Appendix B Error Messages PR4000F 58 This page left blank...
Страница 68: ...Appendix C Default Configuration PR4000F 60 This page left blank...