Chapter Four: Operation
How To Reassemble the Unit After Bakeout
37
Warm Up and Start Up
1. Apply power to the transducer. The power requirement is ±15 VDC (±5%) @ 35 mA.
2. Vacuum-pump the system to a pressure below the transducer’s resolution (0.01% F.S.).
Allow a minimum of 15 minutes for the electronics to stabilize.
3. Zero the output of the transducer by adjusting the zero potentiometer on the electronics
unit or on the power supply/readout unit (if using an MKS power supply/readout unit).
The instrument is ready for normal operation.