How To Prepare for Maximum Bakeout
Chapter Four: Operation
34
How To Prepare for Maximum Bakeout
The Type 121 sensor is bakeable up to 200° C with the sensor housing removed and the cable
disconnected. The following tools are required to remove the sensor housing:
•
Flat blade screwdriver
•
Philips screwdriver
•
¼” nut driver
•
Needle-nose pliers
Refer to Figure 5 when removing or replacing the sensor housing.
Face Plate
Lower Standoff
(2 places)
Coaxial Cable
(white)
Cable Clamping
Assembly
Upper Standoff
(2 places)
Black Shrink
Tubing
Cable Strain Relief Screw
(2 places)
Housing Screw Top
(2 places)
Coaxial Cable
(orange)
Middle Standoff
(2 places)
Pin Connector
(2 places)
Housing Screw Bottom
(2 places)
Figure 5: Sensor Assembly
Содержание Baratron 121A
Страница 3: ...113733 P1 Rev C 7 98 MKS Baratron Type 121A Absolute Pressure Transducer...
Страница 7: ...Table of Contents v Appendix D High Pressure Fitting Selection 49 Description 49 Index 51...
Страница 8: ...Table of Contents vi...
Страница 10: ...List of Figures and Tables viii...
Страница 30: ...Customer Support Chapter One General Information 20 This page intentionally left blank...
Страница 42: ...Labels Chapter Three Overview 32 This page intentionally left blank...
Страница 48: ...How To Reassemble the Unit After Bakeout Chapter Four Operation 38 This page intentionally left blank...
Страница 50: ...Overpressure Considerations Chapter Five Theory of Operation 40 This page intentionally left blank...
Страница 52: ...General Information Chapter Six Maintenance 42 This page intentionally left blank...
Страница 58: ...Companion Products Appendix C Companion Products and Cables 48 This page intentionally left blank...
Страница 60: ...Description Appendix D High Pressure Fitting Selection 50 This page intentionally left blank...
Страница 62: ...Index 52...