Environmental Requirements
Chapter Two: Installation
24
Environmental Requirements
Follow these requirements when installing and using a Type 121 Absolute Pressure Transducer.
1. The normal operating temperature range, for the sensor and cable, is 0° C to 150° C
(32° F to 302° F).
The Type 121 sensor is bakeable to 200° C (392° F) with the sensor housing and cable
removed.
2. The normal operating temperature range, for the electronics unit (signal conditioner), is
0° C to 50° C (32° F to 122° F).
3. The power input required is ±15 VDC (±5%) @ 35 mA.
4. The overpressure limit is 35 psia or 120% of sensor Full Scale, whichever is greater.
5. It is important that the unit be kept free from vibration, because the diaphragm, when not
subject to gas damping at low pressure, may become susceptible to resonance. The
1 Torr F.S. units are the most sensitive; any vibration that does exist should be isolated
from the unit with vibration isolation mounting and a bellows coupling (MKS-supplied
or MKS-compatible).
For additional Type 121 transducer requirements refer to Appendix A: Product Specifications,
page 43.
Содержание Baratron 121A
Страница 3: ...113733 P1 Rev C 7 98 MKS Baratron Type 121A Absolute Pressure Transducer...
Страница 7: ...Table of Contents v Appendix D High Pressure Fitting Selection 49 Description 49 Index 51...
Страница 8: ...Table of Contents vi...
Страница 10: ...List of Figures and Tables viii...
Страница 30: ...Customer Support Chapter One General Information 20 This page intentionally left blank...
Страница 42: ...Labels Chapter Three Overview 32 This page intentionally left blank...
Страница 48: ...How To Reassemble the Unit After Bakeout Chapter Four Operation 38 This page intentionally left blank...
Страница 50: ...Overpressure Considerations Chapter Five Theory of Operation 40 This page intentionally left blank...
Страница 52: ...General Information Chapter Six Maintenance 42 This page intentionally left blank...
Страница 58: ...Companion Products Appendix C Companion Products and Cables 48 This page intentionally left blank...
Страница 60: ...Description Appendix D High Pressure Fitting Selection 50 This page intentionally left blank...
Страница 62: ...Index 52...