Pressure Connections
Chapter Two: Installation
28
Pressure Connections
Standard Fitting
The standard tubing is ½ inch (12.7 mm). To connect the sensor to a Swagelok Ultra-Torr
®
compression fitting, follow the steps below:
1. Be sure the tubing is clean and free of axial scratches.
2. Insert the tubing through the compression nut and O-ring all the way to the shoulder.
3. Tighten the nut.
Optional Fittings
The following fittings are available as options when specified:
•
NW-16-KF
•
Mini-CF
•
Swagelok 8-VCR (female)
Refer to Appendix D: High Pressure Fitting Selection, page 49, for more information.
Caution
During high-temperature operation, (150° C, or 200° C with sensor
housing and cable removed), use one of the optional metal seal
fittings.
Содержание Baratron 121A
Страница 3: ...113733 P1 Rev C 7 98 MKS Baratron Type 121A Absolute Pressure Transducer...
Страница 7: ...Table of Contents v Appendix D High Pressure Fitting Selection 49 Description 49 Index 51...
Страница 8: ...Table of Contents vi...
Страница 10: ...List of Figures and Tables viii...
Страница 30: ...Customer Support Chapter One General Information 20 This page intentionally left blank...
Страница 42: ...Labels Chapter Three Overview 32 This page intentionally left blank...
Страница 48: ...How To Reassemble the Unit After Bakeout Chapter Four Operation 38 This page intentionally left blank...
Страница 50: ...Overpressure Considerations Chapter Five Theory of Operation 40 This page intentionally left blank...
Страница 52: ...General Information Chapter Six Maintenance 42 This page intentionally left blank...
Страница 58: ...Companion Products Appendix C Companion Products and Cables 48 This page intentionally left blank...
Страница 60: ...Description Appendix D High Pressure Fitting Selection 50 This page intentionally left blank...
Страница 62: ...Index 52...