Chapter Two: Installation
Interface Cables
23
Sensor
The Type 121 transducer comes equipped with an 8 foot overall metal braided shielded cable
attached to the sensor. The other end of the cable has a Type “D” connector which plugs into the
electronics unit.
Note
An optional cable is available up to 20 feet (part number CA-105434-
SPCB). Consult the MKS if you need a longer length cable.
Generic Shielded Cable Guidelines
Should you choose to manufacture your own cables, follow the guidelines listed below:
1. The cable must have an overall metal braided shield, covering all wires. Neither aluminum
foil nor spiral shielding will be as effective; using either may nullify regulatory compliance.
2. The connectors must have a metal case which has direct contact to the cable’s shield on the
whole circumference of the cable. The inductance of a flying lead or wire from the shield to
the connector will seriously degrade the shield’s effectiveness. The shield should be
grounded to the connector before its internal wires exit.
3. With very few exceptions, the connector(s) must make good contact to the device’s case
(ground). “Good contact” is about 0.01 ohms; and the ground should surround all wires.
Contact to ground at just one point may not suffice.
4. For shielded cables with flying leads at one or both ends; it is important at each such end, to
ground the shield before the wires exit. Make this ground with absolute minimum length. (A
¼ inch piece of #22 wire may be undesirably long since it has approximately 5 nH of
inductance, equivalent to 31 ohms at 1000 MHz). After picking up the braid’s ground, keep
wires and braid flat against the case. With very few exceptions, grounded metal covers are
not required over terminal strips. If one is required, it will be stated in the Declaration of
Conformity or in the instruction manual.
5. In selecting the appropriate type and wire size for cables, consider:
A. The voltage ratings.
B. The cumulative I
2
R heating of all the conductors (keep them safely cool).
C. The IR drop of the conductors, so that adequate power or signal voltage gets to the device.
D. The capacitance and inductance of cables which are handling fast signals, (such as data
lines or stepper motor drive cables).
E. That some cables may need internal shielding from specific wires to others; please see the
instruction manual for details regarding this matter.
Содержание Baratron 121A
Страница 3: ...113733 P1 Rev C 7 98 MKS Baratron Type 121A Absolute Pressure Transducer...
Страница 7: ...Table of Contents v Appendix D High Pressure Fitting Selection 49 Description 49 Index 51...
Страница 8: ...Table of Contents vi...
Страница 10: ...List of Figures and Tables viii...
Страница 30: ...Customer Support Chapter One General Information 20 This page intentionally left blank...
Страница 42: ...Labels Chapter Three Overview 32 This page intentionally left blank...
Страница 48: ...How To Reassemble the Unit After Bakeout Chapter Four Operation 38 This page intentionally left blank...
Страница 50: ...Overpressure Considerations Chapter Five Theory of Operation 40 This page intentionally left blank...
Страница 52: ...General Information Chapter Six Maintenance 42 This page intentionally left blank...
Страница 58: ...Companion Products Appendix C Companion Products and Cables 48 This page intentionally left blank...
Страница 60: ...Description Appendix D High Pressure Fitting Selection 50 This page intentionally left blank...
Страница 62: ...Index 52...