Series 943 Cold Cathode Vacuum Sensor
Instruction Manual - #100009887
6.4 Troubleshooting
Table 6-1 General Symptoms/Possible Causes
Symptom
Possible Causes
Remedy
Pressure reading is
inaccurate or erratic
1. Pressure may be above the
2. System's range.
3. External magnetic field may be
interfering with sensor operation.
4. Sensor may need cleaning.
5. Controller may be out of calibration.
6. Gas in system may not be air or
nitrogen
1.
Turn OFF Controller. Do not operate
above 10
-2
Torr.
2.
Move magnetic field source away
from sensor or vice versa.
3.
Rebuild sensor intern
als (see Section
6.7, Accessories).
4.
Return to MKS for repair.
5.
Use a correct conversion factor.
No pressure reading on
display
1.
Controller might not be plugged into
proper power source.
2.
Power switch may be OFF.
3.
Power fuse(s) may be blown.
4. Power supply may be defective
1.
Plug into functional AC outlet with
the correct voltage.
2.
Turn power ON.
3.
Replace fuse.
4.
Return to MKS for repair.
LO is displayed
1.
Ion current or high voltage cable may
be disconnected or may
have a bad or
broken connection.
2.
Sensor discharge may not have yet
started.
3.
Pressure may be below
4.
System's range.
5.
Pressure may be above
System's
range.
1.
Check cable connections.
Replace
cable if necessary.
2.
Temporarily backfill to a higher
pressure (10
-4
Torr).
3.
The Series 943 Controller is
operating normally.
4.
Turn high voltage OFF and set
pressure below 10
-2
Torr.
PRO is displayed
5. Pressure is above the protect set
point.
1.
Change protect set point or reduce
system pressure.
Set point relay will not
operate
1. Set point voltage may be incorrectly
set.
2. Short in or incorrectly wired
Accessory connector.
3. Defective PC board.
1.
Check set point.
2.
Rewire connector.
3.
Return to MKS for repair.
High Voltage
To prevent electrical shock, turn OFF electrical power before
servicing the Module. Do not touch any gauge pins while the
sensor tube is under vacuum.