5
Operation
34
Series 943 Cold Cathode Vacuum Sensor
Instruction Manual - #100009887
2.
Use a small screwdriver to adjust its potentiometer until the reading coincides with the desired
set point pressure.
If the protect set point turns the high voltage OFF, PRO will appear on the display. Turn OFF the
high voltage switch and then back ON again to enable the high voltage.
The protect set point cannot work if the system's pressure is in the sensor's "roll-back" pressure
region (see Figure 5-4).