Operation
30
Series 943 Cold Cathode Vacuum Sensor
Instruction Manual - #100009887
5.3.2 Electrometer
The electrometer performs a logarithmic conversion to measure the sensor tube's current and high
voltage. This electrometer has a dynamic range wide enough to cover the entire pressure range with
no electronic range changing.
5.3.3 Analog Processing
The processing circuitry combines the electrometer signals and scales the result to produce a
pressure dependent voltage signal. This voltage drives the A/D converter, unprocessed analog
output, and the set point relay circuitry.
5.3.4 Set Point Comparators
This part of the circuit compares the output of the signal processing circuitry to the value of the set
point adjusting potentiometers and turns the process control relays on and off accordingly.
5.3.5 Microprocessor
This part of the circuit reads the output of the A/D converter, displays pressure on the LED display,
and produces a pressure dependent voltage signal using a D/A converter. See Figure 5-3.
Figure 5-3: 943 Controller Block Diagram