4
Installation
23
Series 943 Cold Cathode Vacuum Sensor
Instruction Manual - #100009887
Figure 4-3
:
Cold Cathode Analog Output Voltage vs. Pressure
4.1.7 Remote High Voltage Enable/Disable
The cold cathode sensor's high voltage may be turned OFF or ON using a set point from a low
vacuum sensor, such as a Pirani or convection Pirani. Use pins 9 and 10 to connect the Controller
to the sensor for this external control.
To use this input, toggle the high voltage switch to Remote. If the high voltage is disabled by the
protect set point, the Remote connection must be set to first disable and then enable the high
voltage to turn it back ON.