41
4 Specifications
No. 99MBE048B
4.8
Changing the Output Settings
Tips
Output setting change is available only on the scale unit with the main signal output type B (two-phase square
wave + external reset input) or C (two-phase square wave + two-phase sine wave).
You can change the output settings with the DIP switch DPSW2 on the I/F Box board.
1
4
3
8
7
6
2
5
ON
Minimum edge-to-edge interval
Direction switch
Resolution
Alarm output type
Minimum edge-to-edge interval
Minimum edge-to-
edge interval
1
2
Factory setting
125 ns
On
On
250 ns
On
Off
●
500 ns
Off
On
1000 ns
Off
Off
Tips
The factory settings are subject to change upon customer's request.
Resolution
Resolution
3
4
5
6
Factory setting
0.01 μm
Off
On
Off
On
0.02 μm
Off
Off
Off
On
0.05 μm
Off
On
On
Off
0.1 μm
On
Off
On
Off
●
The operation is not guaranteed for any other setting than the above.
Tips
The factory settings are subject to change upon customer's request.