17
3 Installation onto the Machine Main Unit
No. 99MBE048B
3.5
Adjusting the Detector Signals
After mounting the main scale and Detector and connecting the feedback cable, adjust the Detector
signals.
3.5.1 Adjusting the Detector Mounting Position
1
Remove the two screws and remove the I/F Box cover.
2
Set the oscilloscope as follows.
Item
Setting description
Measurement pressure range
0.5 V/div (50 mV/div when using a 10:1 probe)
Scan mode
X-Y
3
Clip the two probes and GND probe of the oscilloscope to Vref and GND on the I/F Box, respectively.
Main signal output type: A (two-phase sine wave)
TP5:GND
TP8:ø Z
TP1: øA
TP2:øB
TP3:Vref
VR3:Zpp
VR2:Zdc
VR1:ø A, B