27
4 Specifications
No. 99MBE048B
4
Specifications
4.1
Specifications
Item
Specification
Detection method
Optical reflection type linear encode
Main scale grid pitch
8 μm
Main signal output pitch
4 μm
Output signal type
A: Two-phase sine wave
B: Two-phase square wave (reset input specification)
C: Two-phase square wave, two-phase sine wave
D: Differential 1 Vpp sine wave
Effective length
10 mm–3000 mm
Indication accuracy (20 °C)
Effective length of 10 mm–300 mm: ±0.5 μm
Effective length of 350 mm–500 mm: ±1.0 μm
Effective length of 600 mm–1000 mm: ±2.0 μm
Effective length of 1100 mm–3000 mm: ±2.0 μm/m
Coefficient of linear expansion
≈8×10
–6
/K
Maximum response speed
1200 mm/s (at two-phase sine wave output)
For a two-phase square wave, the value changes depending on
the resolution as follows (when the minimum edge-to-edge interval
is 125 ns):
0.1 μm: 720 mm/s
0.05 μm: 360 mm/s
0.02 μm: 144 mm/s
0.01 μm: 72 mm/s
* When the minimum edge-to-edge interval doubles, the maximum
response speed halves.
Scale origin
Available (50 mm pitch, center point for effective length of 10 mm–
80 mm)
Response speed at scale origin
detection
20 mm/s
Power supply voltage
DC5 V±5 %
Maximum current consumption A: 120 mA
B: 250 mA
C: 250 mA
D: 190 mA
Used temperature range
0 °C–40 °C
Storage temperature range
-20 °C–60 °C
Used/storage humidity range
20 %RH–80 %RH (non condensation)
Alarm Display function
A scale unit alarm is indicated with a LED on the I/F Box.