8
3.8
CR4000 Instructional Video (CR40x)
The CR4000 system instruction video is available in several formats: CD, NTSC, SECAM or
PAL. A videotape or CD is supplied with each CR4000 system. The format received is customer
specified. The video contains an overview of the CR4000 system and is intended to complement
this manual. It can also be used as an instructional aid.
3.9
Ophthalmic Probes
All probe shafts and tips are stainless steel. Refer to Table II for a listing of the MIRA
Ophthalmic Cryo Probes. The gas expansion chamber is at the very tip of the probes,
thereby assuring that freezing and defrosting occurs only at the target area. Warm high-
pressure gas insulates the shaft during freezing and defrosting cycles eliminating the need
for and insulating boot while reducing the danger of freezing adjacent tissues. The
handles present an unobstructed view of the probe shaft and remain comfortable to the
hand during use.
TABLE II Probe Model Numbers
Model No;
Description
Tip Diameter Ice Ball Size
CR4010 Curved Retinal
2.8mm
3.5mm
CR4015 Straight Retinal
2.8mm
3.5mm
CR4022
Straight Cataract
2mm
3.0mm
CR4023 Curved Cataract
2mm
3.0mm
CR4025 Mini Curved Cataract
1.5mm
2.0mm
CR4030 Curved Glaucoma
3.5.mm
4.0mm
CR4040 Vitreous
1.5.mm
2.5mm
CR4045 Mini Straight Vitreous
1.5 mm
1.5mm
CR4075 Curved Hammerhead
2.8mm x 6mm long
4.5mm
CR4080 Curved Baby Hammerhead
1.5mm x 4mm long
2.5mm
CR4085 Gaydon ROP
2mm
3.0mm
CR4090 Abramson Retinoblastoma
1.5mm
2.0mm
CR3095
Lashkari Trichiasis
2.8mm
2.5mm
4.0
INITIAL SET-UP PROCEDURE
IMPORTANT: To be performed or double checked each time an element of the CR4000
system is added, changed or disassembled.
Obtain a non-siphon gas cylinder of N
2
O or CO
2
. The type of cylinder and gas selected will
determine the connector
needed.
See table 1 for a complete selection of gas cylinder connectors.
WARNING;
Siphon gas cylinders will vent gas in a liquid state when the valve is opened.
Do
not use siphon gas cylinders