Microfiber Filter Assembly (Cont’d) 

O-Rings are available from MIRA or your distributor (CR4350). The filter element 
should be changed every 30 to 45 days depending on use. 

 

3.5 

Microfiber Filter Element and O-Ring replacement procedure (CR4350) 

1. Turn the gas cylinder valve OFF 
2. Depressurize the CR4000 system. Turn “ON/OFF’ switch to an in-between setting. Gas can 
   be heard escaping from the exhaust port. 
3. Unscrew, with a wrench, the bottom cylinder of the Micro-fiber Filter Assembly. 
4. Remove the micro fiber filter element by sliding it off the metal core. 
5. Inspect the O-Ring and replace if needed. 
6. Install a new Micro-fiber Filter element (every 30 to 45 days depending on use). 
7. Screw the bottom cylinder back on hand tight, then tighten no more than 1/4 turn. 

Optional; 

Run test procedures as described in section 6.0. 

 

Sintered metal Filter element and O-Ring replacement procedure (CR5350)

 

Proceed as in items 1, 2 and 3 above 

 
 

4. Remove the filter element 

 

5. Inspect the sealing ring and replace unless it shows signs of damage. 

 

6. Install a new filter element. With normal use this should last up to twelve months. 

 

7. Screw filter cover back in place and tighten by hand. Use a wrench to tighten a further 

 

    

¼

 turn  

 

Optional; 

Run test procedures as described in section 6.0 

 
 

3.6 

25ft Scavenger Hose

 

(CR4160)

 

Connect one end of the 25ft. Scavenger hose to the barbed fitting on the console back panel and 
the  other  end  to  an  aspiration  system  or  equivalent.  The  hose  is  used  to  remove  exhausted  gas 
from the treatment area. 

 

TABLE 1 

 

Gas Cylinder Connectors. 

 

Model No; 

Description

   

 

CR4210 

Hand-tight N

2

O connector for “C” 8in. Cylinder (CGA Thread) 

CR4310 

Hand-tight N

2

O connector/Filter for “C” 8in cylinder (CGA Thread) 

CR4220 

Hand-tight CO

2

 connector for “C” 8in cylinder (CGA Thread) 

CR4320 

Hand-tight N

2

O connector/filter for “C” 8in cylinder (CGA Thread) 

CR4250 

Universal T-Yoke connector for “E” 4in cylinder (not available in US) 

 

 

Содержание CR4000

Страница 1: ...First through innovation 0086 Mira Inc 414 Quaker Highway Uxbridge MA 01569 Tel 508 278 7877 800 847 6472 Fax 508 278 4555 p 1 OPHTHALMIC CRYOUNIT CR4000 OPERATIONAL INSTRUCTION MANUAL...

Страница 2: ...er Hose CR4160 6 3 7 Cryo Equipment Carrier Single Dual Cylinder capacity CR4150 7 3 8 CR4000 Instructional Video CR40x 8 3 9 Ophthalmic Probes 8 4 0 Initial Set Up Procedure 8 5 0 Foot Switch Operati...

Страница 3: ...t which can obscure the surgeon s view The MIRA CR4000 has a rapid freeze and defrost cycle available providing greater control of application time Temperature settings for both CO2 and N2O allow for...

Страница 4: ...or changed When a cylinder is heated above room temperature it may exceed maximum pressure 900psi red zone In this situation allow the cylinder to equilibrate to room temperature Approx 70 F before us...

Страница 5: ...o Carriers that come with factory installed sintered metal filters With a Micro fiber or sintered metal Filter assembly attached For use with basic CR4000 systems IMPORTANT Micro fiber or sintered met...

Страница 6: ...ures as described in section 6 0 Sintered metal Filter element and O Ring replacement procedure CR5350 Proceed as in items 1 2 and 3 above 4 Remove the filter element 5 Inspect the sealing ring and re...

Страница 7: ...r N2O CR4290 T Yoke assembly for E 4in Cylinder CO2 CR4390 T Yoke assembly with Micro filter for E 4in Cylinder CO2 3 7 Cryo Equipment Carrier Single Dual Cylinder capacity CR4150 The CR4150 is specia...

Страница 8: ...ues The handles present an unobstructed view of the probe shaft and remain comfortable to the hand during use TABLE II Probe Model Numbers Model No Description Tip Diameter Ice Ball Size CR4010 Curved...

Страница 9: ...er hose to the rear panel fitting labeled System Exhaust Connect the other end to an aspiration system or equivalent Note Use of the scavenger hose will result in quieter operation of the CR4000 Check...

Страница 10: ...he 5 C temperature setting for Carbon Dioxide CO2 6 3 4 Depress the foot pedal for 10 seconds and then release 6 3 5 Wait 60 seconds while the exhaust gas is escaping from the rear scavenger port IMPO...

Страница 11: ...for repair WARNING Do not alter the shape of the probe tip The shape of the probe tip is vital to its proper function Any change to the probe tip will void the manufacturer s warranty 8 2 Cleaning The...

Страница 12: ...s completely non electric there are no electrical shock hazards The micro fiber filter Sintered metal filter assemblies prevent foreign particles or oils from entering the CR4000 console and probes Re...

Страница 13: ...efore patient treatment Turn the gas cylinder OFF and depressurize the CR4000 system before changing the Micro filter Sintered metal filter or O ring When the probe is not in use the protective cap sh...

Страница 14: ...er Place all probes in plastic bags Enter a brief note that includes the following Return Goods Authorization RGA Your full address for return of shipment Please include specific department Descriptio...

Страница 15: ...ncurred by the buyer with the purchase or operation of the equipment or components thereof Repair to damaged parts will be made at a minimal charge The buyer shall pay all shipping charges on the equi...

Отзывы: