Mira CR4000 Скачать руководство пользователя страница 10

 

 

 

 

 

 First through innovation

 

 

 

 

0086

 

 

Mira, Inc.*414 Quaker Highway * Uxbridge, MA 01569 * Tel: 508-278-7877 * 800 847-6472 * Fax: 508-278-4555    p.

10

 

                     

6.0 

TEST PROCEDURE 

To be performed prior to patient treatment

 

 

6.1 

Select an Ophthalmic Cryo Probe

 

and unscrew the protective cap

 

from the probe jack. 

 

6.2

 

Insert the probe jack all the way into the receptacle on the front panel. 

IMPORTANT

: When the probe is not in use, the protective cap  should be kept on the 

probe jack to prevent foreign matter and moisture from entering the probe system during 
handling, cleaning and sterilizing. (See section 8.0). 

 

6.3

   

PRECONDITION  PROBES 

- To assure proper probe operation, 

gas  must  be  purged 

from  the  probes  prior  to  their  use.

  To  properly  purge  gas  from  a  probe,  follow  the 

procedure below, 

 

6.3.1

.  Set the front panel valve “ON/OFF” switch to “ON” 

 

6.3.2.

  Verify there is a minimum of 680psi gauge of pressure. Check the pressure gauge 

Indicator, it should be in the green zone. (680-900psi gauge) 

 

6.3.3

.  Select the -25º C temperature setting for Nitrous Oxide (N

2

O) 

Select the -5º C   temperature setting for Carbon Dioxide (CO

2

)

 

 

6.3.4

.  Depress the foot pedal for 10 seconds and then release. 

 

6.3.5

.  Wait 60 seconds while the exhaust gas is escaping from the rear scavenger port. 

IMPORTANT;  

 

 

 

Most of the purging occurs during the last 30 seconds of the purge  

 

 

 

 

cycle. 

6.3.6.

   Repeat steps 6.3.4.and 6.3.5.twice

.

 This will remove internal condensation as well 

as restore ambient temperature to the system. 

 

6.3.7.

  The system is now ready for use. 

 

Should you wish to perform an “ice ball” test, proceed as follows; 

Remove probe tip cap. Submerge the tip into a solution of sterile saline and 
depress the foot pedal for 10seconds, keeping the tip in the solution. After 
ten  seconds,  with  the  foot  pedal  depressed,  remove  the  probe  tip  from  the 
solution. An ice ball of 1.5 to 2.5 mm larger than the diameter of the probe 
tip should have been formed. Release the foot pedal. In about 4 seconds, the 
ice ball should be free to move when tipping the probe. 

 

6.4

.  

To  remove  the  probe,  allow  the  gas  to  dissipate  for  approximately  60  seconds.  This 
relieves  the  gas  pressure  in  the  system.  If  the  probe  must  be  removed  quickly,  turn  the 
“ON/OFF”  valve  to  “OFF”  while  depressing  the  foot-switch.  This  action  depressurizes 
the system instantly. 

 
 

Содержание CR4000

Страница 1: ...First through innovation 0086 Mira Inc 414 Quaker Highway Uxbridge MA 01569 Tel 508 278 7877 800 847 6472 Fax 508 278 4555 p 1 OPHTHALMIC CRYOUNIT CR4000 OPERATIONAL INSTRUCTION MANUAL...

Страница 2: ...er Hose CR4160 6 3 7 Cryo Equipment Carrier Single Dual Cylinder capacity CR4150 7 3 8 CR4000 Instructional Video CR40x 8 3 9 Ophthalmic Probes 8 4 0 Initial Set Up Procedure 8 5 0 Foot Switch Operati...

Страница 3: ...t which can obscure the surgeon s view The MIRA CR4000 has a rapid freeze and defrost cycle available providing greater control of application time Temperature settings for both CO2 and N2O allow for...

Страница 4: ...or changed When a cylinder is heated above room temperature it may exceed maximum pressure 900psi red zone In this situation allow the cylinder to equilibrate to room temperature Approx 70 F before us...

Страница 5: ...o Carriers that come with factory installed sintered metal filters With a Micro fiber or sintered metal Filter assembly attached For use with basic CR4000 systems IMPORTANT Micro fiber or sintered met...

Страница 6: ...ures as described in section 6 0 Sintered metal Filter element and O Ring replacement procedure CR5350 Proceed as in items 1 2 and 3 above 4 Remove the filter element 5 Inspect the sealing ring and re...

Страница 7: ...r N2O CR4290 T Yoke assembly for E 4in Cylinder CO2 CR4390 T Yoke assembly with Micro filter for E 4in Cylinder CO2 3 7 Cryo Equipment Carrier Single Dual Cylinder capacity CR4150 The CR4150 is specia...

Страница 8: ...ues The handles present an unobstructed view of the probe shaft and remain comfortable to the hand during use TABLE II Probe Model Numbers Model No Description Tip Diameter Ice Ball Size CR4010 Curved...

Страница 9: ...er hose to the rear panel fitting labeled System Exhaust Connect the other end to an aspiration system or equivalent Note Use of the scavenger hose will result in quieter operation of the CR4000 Check...

Страница 10: ...he 5 C temperature setting for Carbon Dioxide CO2 6 3 4 Depress the foot pedal for 10 seconds and then release 6 3 5 Wait 60 seconds while the exhaust gas is escaping from the rear scavenger port IMPO...

Страница 11: ...for repair WARNING Do not alter the shape of the probe tip The shape of the probe tip is vital to its proper function Any change to the probe tip will void the manufacturer s warranty 8 2 Cleaning The...

Страница 12: ...s completely non electric there are no electrical shock hazards The micro fiber filter Sintered metal filter assemblies prevent foreign particles or oils from entering the CR4000 console and probes Re...

Страница 13: ...efore patient treatment Turn the gas cylinder OFF and depressurize the CR4000 system before changing the Micro filter Sintered metal filter or O ring When the probe is not in use the protective cap sh...

Страница 14: ...er Place all probes in plastic bags Enter a brief note that includes the following Return Goods Authorization RGA Your full address for return of shipment Please include specific department Descriptio...

Страница 15: ...ncurred by the buyer with the purchase or operation of the equipment or components thereof Repair to damaged parts will be made at a minimal charge The buyer shall pay all shipping charges on the equi...

Отзывы: