Mira CR4000 Скачать руководство пользователя страница 13

 

 

 

 

 

 First through innovation

 

 

 

 

0086

 

 

Mira, Inc.*414 Quaker Highway * Uxbridge, MA 01569 * Tel: 508-278-7877 * 800 847-6472 * Fax: 508-278-4555    p.

13

 

                     

10.2 

General Safety Precautions

 

 

Review this manual with all personnel involved with the use of the CR4000. 

 

Always mark the empty gas cylinder so it can be replaced immediately 

 

Liquid N

2

O or CO

2

 will damage the CR4000 surgical unit. Use only siphon gas cylinders. 

 

If for any reason the Ophthalmic Cryo CR4000 console, probes or accessories has to be 
returned to MIRA Inc., see section 11.0 for shipping information. 

 

The initial set-up procedure described in section 4.0 should be performed/triple checked 
each time an element of the CR4000 system is added, changed or disassembled. 

 

The test procedure described in section 6.0 should be performed before patient treatment. 

 

Turn  the  gas  cylinder  OFF  and  depressurize  the  CR4000  system  before  changing  the 
Micro-filter/Sintered metal filter or “O” ring 

 

When  the  probe  is  not  in  use,  the  protective  cap  should  be  kept  on  the  probe  jack  to 
prevent  foreign  matter  and  moisture  from  entering  the  probe  system  during  storage, 
handling, cleaning and sterilization. 

 

 

 

Also,  please  ensure  that  the  probe  tip  cap  is  in  place  whenever  the  probe  is  not  in 
use.  Do  not  change  the  shape  of  the  probe  tip  as  this  is  vital  to  the  proper 
functioning of the device. 

 
To assure proper probe operation, gas must be purged from the probes prior to their use. 

 

The  probes  can  be  cleaned  in  accordance  with  protocols  (see  addendum  A).  The  entire 
probe should never be submerged in any solution. 

 

DO NOT

 expose the gas cylinder to direct heat. 

 

Improper storage temperatures or overfilling cylinders can cause equipment malfunction.   

 

Non-siphon cylinders must be used and should be stored in the upright position.   
 
 
 
 
 
 
 
 

Содержание CR4000

Страница 1: ...First through innovation 0086 Mira Inc 414 Quaker Highway Uxbridge MA 01569 Tel 508 278 7877 800 847 6472 Fax 508 278 4555 p 1 OPHTHALMIC CRYOUNIT CR4000 OPERATIONAL INSTRUCTION MANUAL...

Страница 2: ...er Hose CR4160 6 3 7 Cryo Equipment Carrier Single Dual Cylinder capacity CR4150 7 3 8 CR4000 Instructional Video CR40x 8 3 9 Ophthalmic Probes 8 4 0 Initial Set Up Procedure 8 5 0 Foot Switch Operati...

Страница 3: ...t which can obscure the surgeon s view The MIRA CR4000 has a rapid freeze and defrost cycle available providing greater control of application time Temperature settings for both CO2 and N2O allow for...

Страница 4: ...or changed When a cylinder is heated above room temperature it may exceed maximum pressure 900psi red zone In this situation allow the cylinder to equilibrate to room temperature Approx 70 F before us...

Страница 5: ...o Carriers that come with factory installed sintered metal filters With a Micro fiber or sintered metal Filter assembly attached For use with basic CR4000 systems IMPORTANT Micro fiber or sintered met...

Страница 6: ...ures as described in section 6 0 Sintered metal Filter element and O Ring replacement procedure CR5350 Proceed as in items 1 2 and 3 above 4 Remove the filter element 5 Inspect the sealing ring and re...

Страница 7: ...r N2O CR4290 T Yoke assembly for E 4in Cylinder CO2 CR4390 T Yoke assembly with Micro filter for E 4in Cylinder CO2 3 7 Cryo Equipment Carrier Single Dual Cylinder capacity CR4150 The CR4150 is specia...

Страница 8: ...ues The handles present an unobstructed view of the probe shaft and remain comfortable to the hand during use TABLE II Probe Model Numbers Model No Description Tip Diameter Ice Ball Size CR4010 Curved...

Страница 9: ...er hose to the rear panel fitting labeled System Exhaust Connect the other end to an aspiration system or equivalent Note Use of the scavenger hose will result in quieter operation of the CR4000 Check...

Страница 10: ...he 5 C temperature setting for Carbon Dioxide CO2 6 3 4 Depress the foot pedal for 10 seconds and then release 6 3 5 Wait 60 seconds while the exhaust gas is escaping from the rear scavenger port IMPO...

Страница 11: ...for repair WARNING Do not alter the shape of the probe tip The shape of the probe tip is vital to its proper function Any change to the probe tip will void the manufacturer s warranty 8 2 Cleaning The...

Страница 12: ...s completely non electric there are no electrical shock hazards The micro fiber filter Sintered metal filter assemblies prevent foreign particles or oils from entering the CR4000 console and probes Re...

Страница 13: ...efore patient treatment Turn the gas cylinder OFF and depressurize the CR4000 system before changing the Micro filter Sintered metal filter or O ring When the probe is not in use the protective cap sh...

Страница 14: ...er Place all probes in plastic bags Enter a brief note that includes the following Return Goods Authorization RGA Your full address for return of shipment Please include specific department Descriptio...

Страница 15: ...ncurred by the buyer with the purchase or operation of the equipment or components thereof Repair to damaged parts will be made at a minimal charge The buyer shall pay all shipping charges on the equi...

Отзывы: