Mira CR4000 Скачать руководство пользователя страница 3

 

 

 

 

 

 First through innovation

 

 

 

 

0086

 

 

Mira, Inc.*414 Quaker Highway * Uxbridge, MA 01569 * Tel: 508-278-7877 * 800 847-6472 * Fax: 508-278-4555    p.

3

 

                     

 

1.0 INTRODUCTION

 

 

This manual should be reviewed with all personnel involved with CR4000 systems to insure proper and 
safe use. 
The  MIRA  CR4000  and  the  interchangeable  probes  were  designed  to  meet  the  exacting  and  practical 
needs of Ophthalmologists who perform Cryosurgery. The system rapidly freezes or defrosts probe tips 
while  the  ergonomically  designed  handles  remain  warm.  It  is  non-electric,  can  be  used  with  medical 
grade  CO

2

  or  N

2

O  (non-siphon  cylinders)  and  has  three  pre-selected  temperature  settings.  The  unique 

design of the MIRA Cryo probe isolates the gas expansion chamber at the very tip of the probe, thereby 
assuring that freezing and defrosting the probe, thereby assuring that freezing and defrosting occurs only 
at the target area. Tip-only freeze also eliminates the need for an insulating boot, which can obscure the 
surgeon’s view 
 

The  MIRA  CR4000  has  a  rapid  freeze  and  defrost  cycle  available,  providing  greater  control  of 
application time.  

 
Temperature settings for both CO

2

 and N

2

O allow for pre-selected tip temperatures.  

The choices of temperatures are -65, -35 and -5 (+/-5deg.C) for CO

2

 and -85, -55 and -25 (+/- 5 C) for 

N

2

 

2.0 SYSTEM COMPONENTS 

See section 3 for expanded descriptions. Contact your distributor or MIRA, Inc.  
for more information. 
 
2.1 

Base system   
The MIRA CR4000 Ophthalmic Cryo System includes the Console, the Footswitch pedal 
and the following: 
 
CR40xx 

-CR4000 instruction video (specify NTSC.,PAL or SECAM) 

CR4160 

-25ft scavenger hose for system exhaust 

CR4260 

-60in. high-pressure hose for gas input 

CR43xx           -12 gas cylinder connector (section 3.3) 
CR4350 

-12 Replacement filters and  6 "O" rings 

 

2.2 

Accessories 
CR40xx 

-CR4000 instruction video (specify NTSC, PAL or SECAM) 

CR4150 

-Dual Cylinder Cryo Carrier with switch-over valve, fittings, hoses and -- 
-filter assembly 

 

CR4270 

-Protective Probe Tip Cap for CR40xx (specify probe type) 

CR4360 

-30in. pressure hose (3) 

CR4400 

-Micro-fiber Filter Assembly  

 

 

CR5400 

-Sintered metal filter pack (3 plus 1 washer) 

 

 

 

 

 

 

 

 

 

 

 

 

 

Содержание CR4000

Страница 1: ...First through innovation 0086 Mira Inc 414 Quaker Highway Uxbridge MA 01569 Tel 508 278 7877 800 847 6472 Fax 508 278 4555 p 1 OPHTHALMIC CRYOUNIT CR4000 OPERATIONAL INSTRUCTION MANUAL...

Страница 2: ...er Hose CR4160 6 3 7 Cryo Equipment Carrier Single Dual Cylinder capacity CR4150 7 3 8 CR4000 Instructional Video CR40x 8 3 9 Ophthalmic Probes 8 4 0 Initial Set Up Procedure 8 5 0 Foot Switch Operati...

Страница 3: ...t which can obscure the surgeon s view The MIRA CR4000 has a rapid freeze and defrost cycle available providing greater control of application time Temperature settings for both CO2 and N2O allow for...

Страница 4: ...or changed When a cylinder is heated above room temperature it may exceed maximum pressure 900psi red zone In this situation allow the cylinder to equilibrate to room temperature Approx 70 F before us...

Страница 5: ...o Carriers that come with factory installed sintered metal filters With a Micro fiber or sintered metal Filter assembly attached For use with basic CR4000 systems IMPORTANT Micro fiber or sintered met...

Страница 6: ...ures as described in section 6 0 Sintered metal Filter element and O Ring replacement procedure CR5350 Proceed as in items 1 2 and 3 above 4 Remove the filter element 5 Inspect the sealing ring and re...

Страница 7: ...r N2O CR4290 T Yoke assembly for E 4in Cylinder CO2 CR4390 T Yoke assembly with Micro filter for E 4in Cylinder CO2 3 7 Cryo Equipment Carrier Single Dual Cylinder capacity CR4150 The CR4150 is specia...

Страница 8: ...ues The handles present an unobstructed view of the probe shaft and remain comfortable to the hand during use TABLE II Probe Model Numbers Model No Description Tip Diameter Ice Ball Size CR4010 Curved...

Страница 9: ...er hose to the rear panel fitting labeled System Exhaust Connect the other end to an aspiration system or equivalent Note Use of the scavenger hose will result in quieter operation of the CR4000 Check...

Страница 10: ...he 5 C temperature setting for Carbon Dioxide CO2 6 3 4 Depress the foot pedal for 10 seconds and then release 6 3 5 Wait 60 seconds while the exhaust gas is escaping from the rear scavenger port IMPO...

Страница 11: ...for repair WARNING Do not alter the shape of the probe tip The shape of the probe tip is vital to its proper function Any change to the probe tip will void the manufacturer s warranty 8 2 Cleaning The...

Страница 12: ...s completely non electric there are no electrical shock hazards The micro fiber filter Sintered metal filter assemblies prevent foreign particles or oils from entering the CR4000 console and probes Re...

Страница 13: ...efore patient treatment Turn the gas cylinder OFF and depressurize the CR4000 system before changing the Micro filter Sintered metal filter or O ring When the probe is not in use the protective cap sh...

Страница 14: ...er Place all probes in plastic bags Enter a brief note that includes the following Return Goods Authorization RGA Your full address for return of shipment Please include specific department Descriptio...

Страница 15: ...ncurred by the buyer with the purchase or operation of the equipment or components thereof Repair to damaged parts will be made at a minimal charge The buyer shall pay all shipping charges on the equi...