5.11
APPENDIX K:
SHORT CUT KEYES AND HIDDEN FUNCTIONS
5.11.1
HIDDEN SHORTCUT KEYS
Below is the list of the most common shortcut keys currently (v.3.35) in use in the EPA3 monitor program.
Note that these are undocumented features and susceptible to change at any time. Caution should be used
when using them, understanding the full implications first.
When Available?
How to Activate?
What Operation Provided?
During Power-Up
Keep any key pressed
(v.3.35 or earlier)
Will load Default Parameter
values, except for:
- Date and Time
- Password
- Stack Buffer Block
- User Cal. Value
- See comment #1
During "MIP 3086…"- display
Up & Down pressed for a few
seconds
Will show the current ‘This is Zero’
Password and let user modify it.
During any Diagnostic Test
Up pressed
Will freeze on that test until up
pressed again.
During any Diagnostic Test after
Communications Check
Down pressed
Will skip the remaining tests
During Main mode
Up pressed (>2secs)
Toggles between the Display
Modes
During Main mode
Down pressed (>2secs)
Toggles between the
Memory/Running Averages
See comment #2
During Status mode with "No
(more) faults" - Display
Next key pressed
Will cycle through Stack Buffer
Block -values
During "Password?"-Display in
New Zero
Push "1111"
Will perform Dirty Stack Zero -
Operation
During "Password?"-Display in
New Zero
Push "2222"
Access to User Cal.-value
modification.
During "Service Mode?" -
Display
Next Pressed (>4s)
Access to Diagnostic Counters. Up
& Down will reset the counters.
Counter is time/date stamped.
During "Hardware Tests?" -
Display
Next Pressed (>4s)
Access to Hardware Zero setting.
During "Dynamic Mode?" -
Display
Down Pressed (>2s)
Access to "Hidden" parameters:
- Process On/Off
- Trax Monitor
- Relay Logic
- Manual Outputs