Gravimetric sampling can be used for calculating K (mass coefficient):
𝐾 =
𝑚
𝐷
m = mass concentration in mg/m3 from mass gravimetric sampling result
K = calculated coefficient
D = optical density (recorded during mass gravimetric sampling)
2.2.5.19
Date/Time Set-Up
This mode enables the user to program the proper date and time of day. The date is given in YY/MM/DD
format. Time is in the 24-hour format.
2.2.6
SERVICE MODES
The following modes detail the menus available by answering Yes to the ‘Service Mode Y/N’ prompt.
All opacity outputs, relays and indicators are frozen during this mode.
The service mode is entered by simultaneously pushing the up and down arrow for > 3 seconds. Use caution
when pushing the up and down buttons. A momentary blink of the display changes the unit from 6-minute
block averages to running averages. The blink will not be too obvious. If this occurs, push the down button
again.
These modes enable the user to perform special tasks to access service values and sub-modes that are
configurable to the users desires or needs. Each sub-mode is prefaced by a yes/no question; the up arrow
selects yes the down arrow no. Upon completing the mode changes, the next key exits the sub-mode; the
status button escapes the service mode. Special key- strokes are outlined in each mode description.