1.
Loosen 1.5 mm set screw on transmitter side. Set screw is in a hole on the right side of the optical
block near the top right.
2.
Remove zero pipe, unclipping it from all the hold down clips.
3.
Remove two signal wires from terminal block XA1. They are the ones labelled ‘main detector’. Note
which colour belongs on which terminal. The label serves as a reminder.
4.
Loosen seal-tight fitting and remove conduit bundle with wires and zero pipe inside.
5.
Close the shutter on the transmitter by leaving it in the automatic position, and opening the
transmitter door.
5.7.1.2
Receiver