Manufacturer’s declaration-electromagnetic immunity
The N507
is intended for use in the electromagnetic environment (for home healthcare) specified
below.
The customer or the user of the N507 should assure that is used in such and environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment-guidance
(for home healthcare environment)
Conducted RF
3 Vrms:
3 Vrms:
Portable and mobile RF communications
equipment should be used no closer to any
part of the N507 including cables, than the
recommended separation distance
calculated from the equation applicable to
the frequency of the transmitter.
Recommended separation distance:
d
= 1,2 √
P
d
= 1,2
√
P
80 MHz to 800 MHz
d
= 2,3
√
P
800 MHz to 2,5 GHz
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in metres (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic
site survey,
a
should be less than the
compliance level in each frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following
symbol:
IEC 61000-4-6
0,15 MHz – 80
MHz
0,15 MHz – 80
MHz
6 Vrms:
6 Vrms:
in ISM and
amateur
in ISM and
amateur
radio bands
between
radio bands
between
0,15 MHz and 80
MHz
0,15 MHz and 80
MHz
80 % AM at 1
kHz e)
80 % AM at 1
kHz e)
Radiated RF
10 V/m
10 V/m
IEC 61000-4-3
80 MHz – 2,7
GHz b)
80 MHz – 2,7
GHz
80 % AM at 1
kHz c)
80 % AM at 1
kHz